background image

© 2017-2020 National Instruments. All rights reserved. 

377027B-01

Apr20

Information is subject to change without notice. Refer to the 

NI Trademarks and Logo Guidelines 

at 

ni.com/trademarks

 for more information 

on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents 
covering NI products/technology, refer to the appropriate location: 

Help»Patents

 in your software, the 

patents.txt

 file on your media, or the 

National Instruments Patents Notice

 at 

ni.com/patents

. You can find information about end-user license agreements (EULAs) and third-party 

legal notices in the readme file for your NI product. Refer to the 

Export Compliance Information 

at 

ni.com/legal/export-compliance

 

for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR 
IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. 
Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and 
restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.

Содержание PXIe-4310

Страница 1: ...ce Temperature Changes 6 Accuracy Verification 6 Adjustment 8 Reverification 9 Specifications 9 Worldwide Support and Services 9 Software Calibrating the PXIe 4310 requires the installation of NI DAQm...

Страница 2: ...nd reference information PXIe 4310 Specifications PXIe 4310 specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx NI DAQmx Help Informat...

Страница 3: ...tute using the requirements listed in Table 1 Table 1 Recommended Equipment for PXIe 4310 Verification Equipment Recommended Model Requirements Calibrator Fluke 5700A If this instrument is unavailable...

Страница 4: ...B 4310 10V Analog Signal Names Signal Names Reference Direction Description COM 0 7 Analog Input Isolated Ground These terminals are the reference point for differential analog input measurements AI 0...

Страница 5: ...tures and temperature drift Allow a warm up time of at least 15 minutes to ensure that the PXIe 4310 measurement circuitry is at a stable operating temperature Ensure that the PXI PXI Express chassis...

Страница 6: ...ettings tab Temperature information is displayed in the Self Calibration and External Calibration sections of the Settings tab Accuracy Verification This section provides instructions for verifying th...

Страница 7: ...acquisition task 9 For each value in Table 4 repeat steps 3 through 8 10 Disconnect the calibrator from the device You have finished verifying the analog input specification on your device Note If ver...

Страница 8: ...le referenceVoltage 7 5 Table 4 PXIe 4310 Analog Input Values Range V Test Point 24 Hour Limits 2 Year Limits Min Max Location Value V Lower Limit V Upper Limit V Lower Limit V Upper Limit V 10 10 Pos...

Страница 9: ...tment verify that you have met the Test Conditions before returning your device to NI Refer to Worldwide Support and Services for assistance in returning the device to NI Specifications Refer to the P...

Страница 10: ...at ni com patents You can find information about end user license agreements EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at n...