PXIe-4310 Calibration Procedure

|

© National Instruments

|

7

Table 4 shows all acceptable values for the PXIe-4310. NI recommends that you verify all 
ranges.
3.

Set the calibrator voltage to the test value indicated in Table 4.

4.

Create and configure a DAQmx task using the settings in Table 3. Set the range 
corresponding to the selected test value from Table 4.

5.

Start the acquisition.

6.

Average the voltage values that you acquired.

7.

Compare the resulting average to the upper and lower limits listed in Table 4. If the result 
is between these values, the device passes the test.

8.

Clear the acquisition task.

9.

For each value in Table 4 repeat steps 3 through 8.

10. Disconnect the calibrator from the device.

You have finished verifying the analog input specification on your device.

Note

If verification results are within th

24-Hour Limits

 listed in Table 4 an 

adjustment is not required, but is still recommended. If you do not want to perform 
an adjustment, you can update the calibration date without making any adjustments 
by initializing an external calibration and closing the external calibration.

Table 3.  

AI Voltage Configuration

Configuration

Value

Physical Channels

Dev1/ai0:7

Sample Mode

Finite number of samples.

Sample Clock Rate

100000

Samples per Channel

10000

Measurement Type

Voltage

Units

V

Terminal Configuration

Differential

Sample Timing Type

Sample Clock

Содержание PXIe-4310

Страница 1: ...ce Temperature Changes 6 Accuracy Verification 6 Adjustment 8 Reverification 9 Specifications 9 Worldwide Support and Services 9 Software Calibrating the PXIe 4310 requires the installation of NI DAQm...

Страница 2: ...nd reference information PXIe 4310 Specifications PXIe 4310 specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx NI DAQmx Help Informat...

Страница 3: ...tute using the requirements listed in Table 1 Table 1 Recommended Equipment for PXIe 4310 Verification Equipment Recommended Model Requirements Calibrator Fluke 5700A If this instrument is unavailable...

Страница 4: ...B 4310 10V Analog Signal Names Signal Names Reference Direction Description COM 0 7 Analog Input Isolated Ground These terminals are the reference point for differential analog input measurements AI 0...

Страница 5: ...tures and temperature drift Allow a warm up time of at least 15 minutes to ensure that the PXIe 4310 measurement circuitry is at a stable operating temperature Ensure that the PXI PXI Express chassis...

Страница 6: ...ettings tab Temperature information is displayed in the Self Calibration and External Calibration sections of the Settings tab Accuracy Verification This section provides instructions for verifying th...

Страница 7: ...acquisition task 9 For each value in Table 4 repeat steps 3 through 8 10 Disconnect the calibrator from the device You have finished verifying the analog input specification on your device Note If ver...

Страница 8: ...le referenceVoltage 7 5 Table 4 PXIe 4310 Analog Input Values Range V Test Point 24 Hour Limits 2 Year Limits Min Max Location Value V Lower Limit V Upper Limit V Lower Limit V Upper Limit V 10 10 Pos...

Страница 9: ...tment verify that you have met the Test Conditions before returning your device to NI Refer to Worldwide Support and Services for assistance in returning the device to NI Specifications Refer to the P...

Страница 10: ...at ni com patents You can find information about end user license agreements EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at n...

Отзывы: