MAX lists all modules installed in the chassis. Your default names may vary.

Note

 If you do not see your module listed, press <F5> to refresh the

list of installed modules. If the module is still not listed, power off the

system, ensure the module is correctly installed, and restart.

4. Record the identifier MAX assigns to the hardware. Use this identifier when

programming the PXIe-4163.

5. Self-test the hardware by selecting the item in the configuration tree and

clicking 

Self-Test

 in the MAX toolbar.

The MAX self-test performs a basic verification of hardware resources.

Related tasks

What Should I Do if the PXIe-4163 Doesn't Appear in MAX?

What Should I Do if the PXIe-4163 Fails the Self-Test?

Self-Calibrating the PXIe-4163

Self-calibration adjusts the PXIe-4163 for variations in the module environment.

Perform a complete self-calibration after the first time you install the PXIe-4163.

1. Install the PXIe-4163 and let it warm up for 30 minutes.

Note

 Warm up begins when the PXI Express chassis has been

powered on and the operating system has completely loaded.

2. Self-calibrate the PXIe-4163 by clicking the 

Self-Calibrate

 button in MAX or

calling niDCPower Cal Self Calibrate or niDCPower_CalSelfCalibrate.

The PXIe-4163 modules are externally calibrated at the factory but you should

perform a self-calibration in all of the following situations:

After first installing the PXIe-4163 in a chassis

After any module that is in the same chassis as the PXIe-4163 is installed,

uninstalled, or moved

© National Instruments

13

PXIe-4163

Содержание PXIe-4163

Страница 1: ...PXIe 4163 2022 06 30...

Страница 2: ...4163 6 PXIe 4163 Pinout and LEDs 7 Configuring the PXIe 4163 in MAX 12 Self Calibrating the PXIe 4163 13 Programming the PXIe 4163 14 Troubleshooting 15 What Should I Do if the PXIe 4163 Doesn t Appea...

Страница 3: ...for more information about minimum system requirements recommended system and supported application development environments ADEs Unpacking the Kit Notice To prevent electrostatic discharge ESD from d...

Страница 4: ...irements specified in this guide and chassis documentation Optional Items PXI Slot Blocker Kit NI part number 199198 01 SHDB62M DB62M LL Low Leakage Cable for SMU 62 D Sub Male to 62 D Sub Male 1 m NI...

Страница 5: ...del is intended for use in indoor applications only Related information http www ni com manuals 1 Not all chassis with slot cooling capacity 58 W can achieve this ambient temperature range Refer to PX...

Страница 6: ...e using the edges or the metal bracket 1 Ensure the AC power source is connected to the chassis before installing the PXIe 4163 The AC power cord grounds the chassis and protects it from electrical da...

Страница 7: ...e into the slot until it is fully inserted 10 Latch the module in place by pulling up on the ejector handle 11 Secure the module front panel to the chassis using the front panel mounting screws Note T...

Страница 8: ...tor pins channels and LEDs A B PXIe 4163 OUTPUT 24 CH Precision SMU ACCES S VOLTAGE ALL CHANNEL SIGNALS 24V MAX 60V MAX to Item A LED Access Status indicator B LED Voltage Status Indicator Table 1 LED...

Страница 9: ...CH14 Output HI CH14 Sense HI CH0 23 Output LO CH9 Output HI CH9 Sense HI CH8 Output HI CH8 Sense HI CH3 Sense HI CH3 Output HI CH0 Output HI CH0 Sense HI CH5 Output HI CH5 Sense HI CH6 Output HI CH6 S...

Страница 10: ...9 Output HI 5 59 Sense HI 60 Output HI 0 to 5 40 Sense LO B 6 56 Sense HI 57 Output HI 7 36 Sense HI 37 Output HI 8 14 Sense HI 15 Output HI 9 12 Sense HI 13 Output HI 10 33 Sense HI 34 Output HI 11 5...

Страница 11: ...I 12 to 17 29 Sense LO D 18 45 Sense HI 46 Output HI 19 25 Sense HI 26 Output HI 20 3 Sense HI 4 Output HI 21 1 Sense HI 2 Output HI 22 22 Sense HI 23 Output HI 23 43 Sense HI 44 Output HI 18 to 23 24...

Страница 12: ...Grouping Channel 8 Grouping Output HI 37 36 35 36 35 34 33 Output HI Sense HI Channel 6 Grouping Sense HI Channel 7 Grouping Output HI Output HI A B C D Bundle Channels A 0 to 5 B 6 to 11 C 12 to 17 D...

Страница 13: ...e 4163 Fails the Self Test Self Calibrating the PXIe 4163 Self calibration adjusts the PXIe 4163 for variations in the module environment Perform a complete self calibration after the first time you i...

Страница 14: ...dio is automatically installed when you install the NI DCPower driver You can access InstrumentStudio in any of the following ways From the Windows start menu select National Instruments Driver Soft F...

Страница 15: ...mples are available from the Start menu in the National Instruments folder C C Available at Program Files IVI Foundation IVI Refer to the Creating an Application with NI DCPower in Microsoft Visual C...

Страница 16: ...Is On Before proceeding verify that the PXIe 4163 appears in MAX If the ACCESS LED fails to light after you power on the chassis a problem may exist with the chassis power rails a hardware module or...

Страница 17: ...tart the system 2 Launch MAX Failed self test 1 Perform self calibration then perform the self test again The PXIe 4163 must be calibrated to pass the self test Failed self calibration 1 Perform self...

Страница 18: ...rvices The NI website is your complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and...

Отзывы: