6. Verify that the module appears in MAX.
7. Reset the module in MAX and perform a self-test.
Related information
■
What Should I Do if the PXIe-4163 Fails the Self-Test?
1. Restart the system.
2. Launch MAX.
■
Failed self-test
1. Perform self-calibration, then perform the self-test again. The
PXIe-4163 must be calibrated to pass the self-test.
■
Failed self-calibration
1. Perform self-calibration again.
3. Power off the chassis.
4. Reinstall the failed module in a different slot.
5. Power on the chassis.
6. Perform the self-test again.
Where To Go Next