1.
In MAX, navigate to the External Calibration section of the Settings tab to update the
Calibration Due Date entry.
2.
Alternatively, use the Update Calibration VI in the NI System Configuration API to set
the calibration due date for either a specific date or an interval in months.
Revision History
Revision
Section
Changes
374877B-01
August 2020
Test Equipment
Updated parameters of digital multimeter
(DMM) and updated PXIe-4135 Calibration
Accessories Kit information.
Verifying Current
Offset Null
Added Verifying Current Offset Null task.
Verifying Current
Offset
Updated as-left offset test limits and added
instructions to reset PLCs.