6.
Verify that the recorded value falls within the test limits.
7.
If more than one limit range is specified, repeat the previous steps using the values
specified in each limit range.
8.
If more than one level range is specified, repeat the previous steps using the values
specified in each level range.
Verifying Load Regulation
Note
Although load regulation is listed as a typical specification for the
PXIe-4135, verification is required to ensure the procedures listed for verifying other
specifications are correct. If the PXIe-4135 fails the load regulation verification
procedure, discontinue use of the device and contact an authorized NI service
representative to request a Return Material Authorization (RMA).
Refer to the following table as you complete the following steps:
Table 7. Load Regulation Verification
Level Range
Limit Range and Limit
Test Point
As-Found/As-Left Limit
10 mA
600 mV
10 mA
2.25 mV
1.
Make the necessary connections for this procedure, as shown in the following figure:
Figure 5. Voltage Measurement Load Regulation
NI-DCPower Device
Sense LO
Sense HI
(triax)
HI
(triax)
LO
CHS GND
Output Shorting Cable Assembly
2.
Set the
niDCPower Output Function
property or
NIDCPOWER_OUTPUT_FUNCTION
attribute to DC Current for the PXIe-4135.
3.
Set the
niDCPower Sense
property or
NIDCPOWER_ATTR_SENSE
attribute to Local.
4.
Set the first specified level range, limit range, and limit on the PXIe-4135.
14
|
ni.com
|
PXIe-4135 Calibration Procedure