background image

NI PXIe-1086 Getting Started Guide

|

© National Instruments

|

11

Cleaning

Cleaning procedures consist of exterior and interior cleaning of the chassis. Refer to your 
module user documentation for information on cleaning the individual CompactPCI or 
PXI Express modules.

Caution

Always

 disconnect all power cables before cleaning or servicing the 

chassis.

Interior Cleaning

Use a dry, low-velocity stream of air to clean the interior of the chassis. Use a soft-bristle brush 
for cleaning around components.

Exterior Cleaning

Clean the exterior surfaces of the chassis with a dry lint-free cloth or a soft-bristle brush. If any 
dirt remains, wipe with a cloth moistened in a mild soap solution. Remove any soap residue by 
wiping with a cloth moistened with clear water. Do not use abrasive compounds on any part of 
the chassis.

Caution

Avoid getting moisture inside the chassis during exterior cleaning, 

especially through the top vents. Use just enough moisture to dampen the cloth.

Caution

Do 

not

 wash the front- or rear-panel connectors or switches. Cover these 

components while cleaning the chassis.

Caution

Do 

not

 use harsh chemical cleaning agents; they may damage the chassis. 

Avoid chemicals that contain benzene, toluene, xylene, acetone, or similar solvents.

Replacing a Modular Power Supply

This section describes how to remove and install a modular power supply for the NI PXIe-1086 
chassis.

Caution

Do not use a power supply from another chassis. Doing so may damage 

your chassis and the power supply.

Removal

The NI PXIe-1086 power supply (part number 782106-01) is a replacement part for the 
NI PXIe-1086 chassis. Before attempting to replace a power supply, verify that there is adequate 
clearance behind the chassis.

The power supplies for this chassis are redundant and hot swappable. If both power supplies are 
installed and functional, you can remove either without disconnecting main AC power from the 
system. If both power supplies are installed, and one has failed, you can remove the failed supply 
without disconnecting main AC power from the system.

Содержание PXIe-1086

Страница 1: ...the Controller 7 Step 3 Install Modules 9 Step 4 Configure the System In Software Using MAX 10 Step 5 Connect Signals to Modules 10 Where To Go Next 10 Maintenance 10 Service Interval 10 Cleaning 11...

Страница 2: ...ircuits Operating personnel and service personnel must not remove protective covers when operating or servicing the chassis Adjustments and service to internal components must be undertaken by qualifi...

Страница 3: ...rom short circuit and overload with automatic recovery Efficiency 70 typical Power disconnect The AC power cables provide main power disconnect Do not position the equipment so that it is difficult to...

Страница 4: ...PXI Filler Panels Optional 9 Removable Feet 10 Fan Door Latch 11 PXI Express Hybrid Peripheral Slots 16x 12 PXI Express PXI Express System Timing Slot 13 PXI Express System Controller Slot 14 Ethernet...

Страница 5: ...ust be set to High to meet the maximum altitude specification Pollution Degree 2 For indoor use only Note Refer to the NI PXIe 1086 User Manual for full specifications Installing the PXI Express Syste...

Страница 6: ...ted spec the power switch LED blinks green Note If two system fans or both power supply fans fail the chassis shuts down automatically protecting the chassis and modules from damage due to overheating...

Страница 7: ...cord grounds the chassis and protects it from electrical damage while you install the controller Caution Refer to your PXI Express system controller user manual for specific instructions and warnings...

Страница 8: ...ler installed in the system controller slot of a chassis Figure 5 PXI Express System Controller Installed in an NI PXIe 1086 Chassis 4 Connect the keyboard mouse and monitor to the appropriate connect...

Страница 9: ...chassis making sure that the injector ejector handle is pushed down Figure 6 Installing PXI PXI Express or CompactPCI Peripheral Modules 5 When you begin to feel resistance push up on the injector ej...

Страница 10: ...abeled PXI System Unidentified Right click this entry to display the pop up menu then select the appropriate system controller model from the Identify As submenu 5 Click the PXI system controller The...

Страница 11: ...ts Use just enough moisture to dampen the cloth Caution Do not wash the front or rear panel connectors or switches Cover these components while cleaning the chassis Caution Do not use harsh chemical c...

Страница 12: ...Before handling the power supply allow the fan to stop spinning 1 Disengage the two captive screws on the rear of the power supply with a flat blade screwdriver 2 Extend the collapsible handle and pu...

Страница 13: ...remove the power drawer 1 Loosen the drawer lever captive screws with a flat blade screwdriver until the threads disengage from the chassis frame 2 Rotate the drawer levers to eject the drawer from th...

Страница 14: ...power drawer from the chassis The power drawer is shown in Figure 8 Note If you are using the NI PXIe 1086 and NI SC Express modules with front mounting terminal blocks together you must remove the SC...

Страница 15: ...our National Instruments product Product registration facilitates technical support and ensures that you receive important information updates from NI National Instruments corporate headquarters is lo...

Страница 16: ...s You can find information about end user license agreements EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal expor...