Appendix A
Specifications
NI 951x User Manual
A-8
ni.com
Online Product Certification
Refer to the product Declaration of Conformity (DoC) for additional
regulatory compliance information. To obtain product certifications and the
DoC for this product, visit
ni.com/certification
, search by module
number or product line, and click the appropriate link in the Certification
column.
Shock and Vibration
To meet these specifications, you must panel mount the system.
Operating vibration
Random (IEC 60068-2-64)..............5 g
rms
, 10 to 500 Hz
Sinusoidal (IEC 60068-2-6) ............5 g, 10 to 500 Hz
Operating shock (IEC 60068-2-27) ........30 g, 11 ms half sine,
50 g, 3 ms half sine,
18 shocks at 6 orientations
Environmental
National Instruments C Series modules are intended for indoor use only,
but may be used outdoors if installed in a suitable enclosure. Refer to the
manual for the chassis you are using for more information about meeting
these specifications.
Operating temperature
(IEC 60068-2-1, IEC 60068-2-2) ...........–40 to 70 °C
Storage temperature
(IEC 60068-2-1, IEC 60068-2-2) ...........–40 to 85 °C
Ingress protection ...................................IP 40
Operating humidity (IEC 60068-2-56) ....10 to 90% RH, noncondensing
Storage humidity (IEC 60068-2-56).......5 to 95% RH, noncondensing
Max altitude ............................................2,000 m
Pollution Degree .....................................2