background image

4

|

ni.com

|

NI 9263 Calibration Procedure

Verification

The following performance verification procedure describes the sequence of operation and 
provides test points required to verify the NI 9263. The verification procedure assumes that 
adequate traceable uncertainties are available for the calibration references. 

Accuracy Verification

Complete the following procedure to determine the As-Found status of the NI 9263.
1.

Set the DMM to Standby mode (STBY).

2.

Connect the DMM to the NI 9263 as shown in Figure 1.

Figure 1.  

Accuracy Verification Connections

3.

Set the DMM to read DC voltage in the minimum range required to cover the NI 9263 
output range of ±10 V.

4.

Acquire a sample
a.

Create and configure an AO voltage channel according to Table 2.

b.

Start the task.

c.

Generate a voltage output according to Table 3.

d.

Wait the appropriate time for the DMM measurement to settle.

e.

Read the NI 9263 output voltage measurement from the DMM.

f.

Clear the task.

Table 2.  

NI 9263 AO Voltage Channel Configuration

Input Range

Scaled Units

Terminal Configuration

Min

Max

-10

10

Volts

Single Ended

Table 3.  

NI 9263 Output Configuration

Samples Per 

Channel

Timeout

Data

1

10.0

A Test Point value indicated in Table 4 or Table 5.

NI 926

3

DMM

AO

x

AO

x

+

HI

LO

Содержание NI 9263

Страница 1: ...Initial Setup 3 Verification 4 Accuracy Verification 4 Adjustment 6 Accuracy Adjustment 6 EEPROM Update 6 Reverification 7 Accuracy Under Calibration Conditions 7 Worldwide Support and Services 8 Sof...

Страница 2: ...ctions and Specifications NI 9263 specific information specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx LabVIEW Help LabVIEW progra...

Страница 3: ...NI 9263 board temperature will be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 10 minutes to ensure that the NI 9263 measurement circuitry is...

Страница 4: ...Accuracy Verification Connections 3 Set the DMM to read DC voltage in the minimum range required to cover the NI 9263 output range of 10 V 4 Acquire a sample a Create and configure an AO voltage chann...

Страница 5: ...le was purchased before 2007 then use the Revision D limits Note Test limits in Table 4 are derived using the values in Table 7 Table 4 NI 9263 Revision D Test Limits Range V Test Point 1 Year Limits...

Страница 6: ...C value obtained from the array of recommended calibration voltages f Wait the appropriate amount of time for the DMM measurement to settle g Read the NI 9263 output voltage measurement from the DMM h...

Страница 7: ...e device to NI Accuracy Under Calibration Conditions The values in the following table are based on calibrated scaling coefficients which are stored in the onboard EEPROM The following accuracy table...

Страница 8: ...ur complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance from NI Ap...

Отзывы: