NI 9263 Calibration Procedure

|

© National Instruments

|

5

5.

Compare the DMM measurement to the test limits in Table 4.

6.

Repeat steps through 5 for each test point in Table 4.

7.

Disconnect the DMM from the NI 9263

.

8.

Repeat steps through 7 for each channel on the NI 9263.

Note

The accuracy and test limits of the NI 9263 vary depending on which revision 

of the module you are using. If the module was purchased after 2007 use the 
Revision E limits. If the module was purchased before 2007 then use the Revision D 
limits.

Note

Test limits in Table 4 are derived using the values in Table 7.

Table 4.  

NI 9263 Revision D Test Limits

Range (V)

Test Point

1-Year Limits

Min

Max

Location

Value (V)

Lower

Limit (V)

Upper

Limit (V)

-10.000

10.000

Max

9.500000

9.460540

9.539460

-10.000

10.000

Mid

0.000000

-0.029960

0.029960

-10.000

10.000

Min

-9.500000

-9.539460

-9.460540

Table 5.  

NI 9263 Revision E Analog Test Limits

Range (V)

Test Point

1-Year Limits

Min

Max

Location

Value (V)

Lower

Limit (V)

Upper

Limit (V)

-10.000

10.000

Max

9.500000

9.482770

9.517230

-10.000

10.000

Mid

0.000000

-0.009630

0.009630

-10.000

10.000

Min

-9.500000

-9.517230

-9.482770

Содержание NI 9263

Страница 1: ...Initial Setup 3 Verification 4 Accuracy Verification 4 Adjustment 6 Accuracy Adjustment 6 EEPROM Update 6 Reverification 7 Accuracy Under Calibration Conditions 7 Worldwide Support and Services 8 Sof...

Страница 2: ...ctions and Specifications NI 9263 specific information specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx LabVIEW Help LabVIEW progra...

Страница 3: ...NI 9263 board temperature will be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 10 minutes to ensure that the NI 9263 measurement circuitry is...

Страница 4: ...Accuracy Verification Connections 3 Set the DMM to read DC voltage in the minimum range required to cover the NI 9263 output range of 10 V 4 Acquire a sample a Create and configure an AO voltage chann...

Страница 5: ...le was purchased before 2007 then use the Revision D limits Note Test limits in Table 4 are derived using the values in Table 7 Table 4 NI 9263 Revision D Test Limits Range V Test Point 1 Year Limits...

Страница 6: ...C value obtained from the array of recommended calibration voltages f Wait the appropriate amount of time for the DMM measurement to settle g Read the NI 9263 output voltage measurement from the DMM h...

Страница 7: ...e device to NI Accuracy Under Calibration Conditions The values in the following table are based on calibrated scaling coefficients which are stored in the onboard EEPROM The following accuracy table...

Страница 8: ...ur complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance from NI Ap...

Отзывы: