background image

Figure 1. NI 9234 Pinout

AI3+

AI3–

AI2+

AI2–

AI0+

AI0–

AI1+

AI1–

Tip

 To minimize ground noise, prevent the metal shells of the BNC

connectors from coming in contact with each other, the modules, or the

chassis.

Signal Descriptions

Signal Signal Description

AI+

Provides DC excitation (when enabled) and positive input signal connection

AI-

Provides excitation return path and signal ground reference

Table 1. Signal Descriptions

Connecting Floating Differential Signals

You can connect floating differential signals to the NI-9234.

Figure 2. Connecting Floating Differential Signals to the NI-9234

NI 9234

AI+

AI–


Signal

Source

© National Instruments

7

NI-9234 Getting Started

Содержание NI 9234

Страница 1: ...NI 9234 Getting Started 2022 07 06...

Страница 2: ...5 Special Conditions for Marine Applications 5 Preparing the Environment 6 Connecting the NI 9234 6 Signal Descriptions 7 Connecting Floating Differential Signals 7 Connecting Grounded Differential Si...

Страница 3: ...lisation L utilisation du produit de toute autre fa on que celle sp cifi e risque de l endommager et de compromettre la protection de s curit int gr e Safety Voltages Connect only voltages that are wi...

Страница 4: ...tool secured cover door where a minimum protection of at least IP54 is provided Caution For Division 2 and Zone 2 applications connected signals must be within the following limits Capacitance 0 2 F...

Страница 5: ...ated in the product specifications These requirements and limits provide reasonable protection against harmful interference when the product is operated in the intended operational electromagnetic env...

Страница 6: ...signing selecting and installing measurement probes and cables to ensure that the desired EMC performance is attained Preparing the Environment Ensure that the environment in which you are using the N...

Страница 7: ...ignal Description AI Provides DC excitation when enabled and positive input signal connection AI Provides excitation return path and signal ground reference Table 1 Signal Descriptions Connecting Floa...

Страница 8: ...of reading per ohm of AI impedance The common mode bias current causes an error only with grounded sources and is not an issue with floating signal sources For best accuracy use low impedance leads wh...

Страница 9: ...port resources including documentation downloads and troubleshooting and application development self help such as tutorials and examples Visit ni com services to learn about NI service offerings such...