background image

Connecting Grounded Differential Signals

You can connect grounded differential signals to the NI-9234.

Figure 3. Connecting Grounded Differential Signals to the NI-9234

NI 9234

AI+

AI–


Signal

Source

Common

Voltage

Mode

Related concepts

Overvoltage Protection

Common-Mode Bias Current

The NI 9234 uses common-mode bias current to bias the current-limiting diodes

when IEPE current is turned off. When the NI 9234 is using grounded signal sources,

this current causes an error that is dependent on the AI- lead impedance. This error

is approximately 50 ppm of range and 15 ppm of reading per ohm of AI- impedance.

The common-mode bias current causes an error only with grounded sources and is

not an issue with floating signal sources. For best accuracy, use low-impedance

leads when connecting grounded signal sources.

Figure 4. Measurement Error Introduced by Common-Mode Bias Current

NI 9234

AI+

AI–


Signal 

Source

AI– Lead

Impedance

Common

Mode

Bias

Current

IEPE Excitation Current

The NI 9234 can also provide an IEPE excitation current for each channel to measure

ground-referenced or floating IEPE sensors. Typical IEPE sensors have a case that is

electrically isolated from the IEPE electronics, so connecting the sensor to the NI

9234 results in a floating connection even though the case of the sensor is

grounded.

ni.com

8

NI-9234 Getting Started

Содержание NI 9234

Страница 1: ...NI 9234 Getting Started 2022 07 06...

Страница 2: ...5 Special Conditions for Marine Applications 5 Preparing the Environment 6 Connecting the NI 9234 6 Signal Descriptions 7 Connecting Floating Differential Signals 7 Connecting Grounded Differential Si...

Страница 3: ...lisation L utilisation du produit de toute autre fa on que celle sp cifi e risque de l endommager et de compromettre la protection de s curit int gr e Safety Voltages Connect only voltages that are wi...

Страница 4: ...tool secured cover door where a minimum protection of at least IP54 is provided Caution For Division 2 and Zone 2 applications connected signals must be within the following limits Capacitance 0 2 F...

Страница 5: ...ated in the product specifications These requirements and limits provide reasonable protection against harmful interference when the product is operated in the intended operational electromagnetic env...

Страница 6: ...signing selecting and installing measurement probes and cables to ensure that the desired EMC performance is attained Preparing the Environment Ensure that the environment in which you are using the N...

Страница 7: ...ignal Description AI Provides DC excitation when enabled and positive input signal connection AI Provides excitation return path and signal ground reference Table 1 Signal Descriptions Connecting Floa...

Страница 8: ...of reading per ohm of AI impedance The common mode bias current causes an error only with grounded sources and is not an issue with floating signal sources For best accuracy use low impedance leads wh...

Страница 9: ...port resources including documentation downloads and troubleshooting and application development self help such as tutorials and examples Visit ni com services to learn about NI service offerings such...

Отзывы: