NI 9215 Calibration Procedure

|

© National Instruments

|

7

h.

Repeat steps d through g for each calibration voltage in the array.

i.

Close the calibration session.

4.

Disconnect the NI 9215 from the calibrator.

5.

Repeat steps through 4 for each channel on the NI 9215.

EEPROM Update

When an adjustment procedure is completed, the NI 9215 internal calibration memory 
(EEPROM) is immediately updated.

If you do not want to perform an adjustment, you can update the calibration date and onboard 
calibration temperature without making any adjustments by initializing an external calibration, 
setting the C Series calibration temperature,

 

and closing the external calibration.

Re-Verification

Repeat the 

Verification

 section to determine the As-Left status of the device.

Note

If any test fails Re-Verification after performing an adjustment, verify that 

you have met the 

Test Conditions

 before returning your device to NI. Refer to 

Where 

to Go for Support

 for assistance in returning the device to NI.

Accuracy Under Calibration Conditions

The following calibration specifications are valid under the following conditions:

Ambient temperature 23 °C ±5 °C

NI 9215 installed in slot 8 of an NI cDAQ-9178 chassis

Slots 1 through 7 of the NI cDAQ-9178 chassis are empty

Note

The limits listed in Table 4 are derived using the values in Table 6.

Note

For operational specifications, refer to the most recent 

NI 9215 Operating 

Instructions and Specifications

 online at 

ni.com/manuals

.

Table 6.  

NI 9215 Accuracy Under Calibration Conditions

Gain Error

Offset Error

0.04%

1.5 mV

Содержание NI 9215

Страница 1: ...irements 2 Test Equipment 3 Test Conditions 3 Initial Setup 3 Verification 4 Accuracy Verification 4 Adjustment 6 Accuracy Adjustment 6 EEPROM Update 7 Re Verification 7 Accuracy Under Calibration Con...

Страница 2: ...and Specifications NI 9215 specific information specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx LabVIEW Help LabVIEW programming...

Страница 3: ...thermal offsets Maintain an ambient temperature of 23 C 5 C The NI 9215 temperature will be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 10 min...

Страница 4: ...ion Complete the following procedure to determine the As Found status of the NI 9215 1 Set the calibrator to Standby mode STBY 2 Connect the NI 9215 to the calibrator Refer to Figure 1 for a connectio...

Страница 5: ...l according to Table 2 b Configure the AI voltage channel timing according to Table 3 c Start the task d Read samples from the NI 9215 e Average the samples for each channel f Clear the task 1 Connect...

Страница 6: ...ust the NI 9215 a Initialize a calibration session on the NI 9215 The default password is NI b Input the external temperature in degrees Celsius c Call the NI 9215 get C Series adjustment points funct...

Страница 7: ...n Repeat the Verification section to determine the As Left status of the device Note If any test fails Re Verification after performing an adjustment verify that you have met the Test Conditions befor...

Страница 8: ...s in the NI DAQmx Readme Refer to the Export Compliance Information at ni com legal export compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes E...

Отзывы: