Connecting to a Spring-Terminal Connector
What to Use
■
NI-9213 spring-terminal connector
■
0.08 mm
2
to 1.0 mm
2
(28 AWG to 18 AWG) copper conductor wire with 7 mm
(0.28 in.) of insulation stripped from the end
■
Flathead screwdriver with a 2.3 mm x 1.0 mm (0.09 in. x 0.04 in.) blade,
included with the NI-9213
What to Do
Complete the following steps to connect wires to the spring-terminal connector.
1
2
3
1. Insert the screwdriver into a spring clamp activation slot to open the
corresponding connector terminal.
2. Press a wire into the open connector terminal.
3. Remove the screwdriver from the activation slot to clamp the wire into place.
High-Vibration Application Connections
If your application is subject to high vibration, NI recommends that you use the
NI-9940 backshell kit to protect connections to the NI-9213.
© National Instruments
9
NI-9213 Getting Started