NI 6711/6713/6731/6733 Calibration Procedure

4.

To find out the date of the last calibration, call

Get_Cal_Date

, which

is included in the

ni671x.dll

.

CalDate

stores the date when the

device was last calibrated.

5.

Set the calibrator to output a voltage of 5.0 V.

6.

Call

Calibrate_E_Series

with the following parameters set as

indicated:

calOP

set to

ND_EXTERNAL_CALIBRATE

setOfCalConst

set to

ND_USER_EEPROM_AREA

calRefVolts

set to

5.0

Note

If the voltage supplied by the source does not maintain a steady 5.0 V, you receive

an error.

7.

Call

Copy_Const

to copy the new calibration constants to the

factory-protected portion of the EEPROM. This function also updates
the calibration date.

8.

Disconnect the calibrator from the device.

The device is now adjusted with respect to the external source. After
the device is adjusted, you can verify the AO operation by repeating the

Verifying Analog Output

section.

Specifications

The following tables are accuracy specifications to use when verifying and
adjusting the NI 6711/6713/6731/6733. The tables show the specifications
for 1-year and 24-hour calibration intervals.

Using the Tables

The following definitions describe how to use the specification tables in
this section.

Range

Range refers to the maximum allowable voltage range of an input or output
signal. For example, if a device is configured in bipolar mode with a range
of 20 V, the device can sense signals b10 and –10 V.

Polarity

Polarity refers to the positive and negative voltages of the input signal that
can be read. Bipolar means the device can read both positive and negative
voltages. Unipolar means that the device can read only positive voltages.

Содержание NI 6731

Страница 1: ...ed for calibrating NI 6711 6713 6731 6733 devices Note Refer to ni com support calibrat mancal htm for a copy of the ni671xCal dll file What Is Calibration Calibration consists of verifying the measur...

Страница 2: ...en calibrated with respect to an external standard However external variables such as temperature can still affect measurements The new calibration constants are defined with respect to the calibratio...

Страница 3: ...imize connections and test conditions during calibration Keep connections to the NI 6711 6713 6731 6733 short Long cables and wires act as antennae picking up extra noise which can affect measurements...

Страница 4: ...w to configure the device in NI DAQ Refer to the NI 671X 673X User Manual for detailed installation instructions You can install this manual when you install NI DAQ 1 Power down the computer 2 Install...

Страница 5: ...DAQ User Manual for PC Compatibles at ni com manuals These two documents provide detailed information about using NI DAQ The function reference help includes information on the functions in NI DAQ Th...

Страница 6: ...those specified by the calibration procedure 2 To internally calibrate the device call the Calibrate_E_Series function with the following parameters set as indicated calOP set to ND_SELF_CALIBRATE set...

Страница 7: ...ifying the Performance of the Counter This procedure verifies the performance of the counter The NI 6711 6713 6731 6733 devices have only one timebase to verify so you only need to verify counter 0 Be...

Страница 8: ...nnot modify these values which provides a level of security that ensures users do not accidentally access or modify any calibration constants adjusted by the metrology laboratory This step in the cali...

Страница 9: ...om the device The device is now adjusted with respect to the external source After the device is adjusted you can verify the AO operation by repeating the Verifying Analog Output section Specification...

Страница 10: ...r the test point value should fall between the upper and lower limit values These limits are expressed in volts Counters Because you cannot adjust the resolution of the counter timers these values do...

Страница 11: ...Programs National Instruments NI DAQ NI DAQ Help and the NI DAQ User Manual for PC Compatibles at ni com manuals for additional information on the software structure Verifying Analog Output Figure 1 V...

Страница 12: ...CTR_Control deviceNumber 0 ND_RESET GPCTR_Set_Application deviceNumber 0 ND_PULSE_TRAIN_GNR GPCTR_Change_Parameter deviceNumber 0 ND_COUNT_1 2 GPCTR_Change_Parameter deviceNumber 0 ND_COUNT_2 2 Select...

Страница 13: ...djusting the NI 6711 6713 6731 6733 Figure 3 Adjusting the NI 6711 6713 6731 6733 Calibrate_E_Series deviceNumber ND_SELF_CALIBRATE ND_USER_EEPROM_AREA 0 Get_Cal_Date deviceNumber CalDate Calibrate_E_...

Отзывы: