background image

NI 6711/6713/6731/6733 Calibration Procedure

10

ni.com

Test Point

The Test Point is the voltage value that is input or output for verification
purposes. This value is broken down into Location and Value. Location
refers to where the test value fits within the test range. Pos FS refers to
positive full-scale, and Neg FS refers to negative full-scale. Value refers to
the voltage to be verified, and Zero refers to the outputting of zero volts.

24-Hour Ranges

The 24-Hour Range column contains the upper limits and lower limits for
the test point value. If the device has been calibrated in the last 24 hours,
the test point value should fall between the upper and lower limit values.
These limit values are expressed in volts.

1-Year Ranges

The 1-Year Range column contains the upper limits and lower limits for the
test point value. If the device has been calibrated in the last year, the test
point value should fall between the upper and lower limit values. These
limits are expressed in volts.

Counters

Because you cannot adjust the resolution of the counter/timers, these values
do not have a 1-year or 24-hour calibration period. However, the test point
and upper and lower limits are provided for verification purposes.

Table 3.

Analog Output for the NI 6711/6713

Range (V)

Polarity

Test Point

24-Hour Ranges

1-Year Ranges

Location

Value (V)

Lower

Limit (V)

Upper

Limit (V)

Lower

Limit (V)

Upper

Limit (V)

0

Bipolar

Zero

0.0

–0.0059300

0.0059300

–0.0059300

0.0059300

20

Bipolar

Pos FS

9.9900000

9.9822988

9.9977012

9.9818792

9.9981208

20

Bipolar

Neg FS

–9.9900000

–9.9977012

–9.9822988

–9.9981208

–9.9818792

Table 4.

Analog Output for the NI 6731/6733

Range (V)

Polarity

Test Point

24-Hour Ranges

1-Year Ranges

Location

Value (V)

Lower

Limit (V)

Upper

Limit (V)

Lower

Limit (V)

Upper

Limit (V)

0

Bipolar

Zero

0.0

–0.0010270

0.0010270

–0.0010270

0.0010270

20

Bipolar

Pos FS

9.9900000

9.9885335

9.9914665

9.9883636

9.9916364

20

Bipolar

Neg FS

–9.9900000

–9.9914665

–9.9885335

–9.9916364

–9.9883636

Содержание NI 6731

Страница 1: ...ed for calibrating NI 6711 6713 6731 6733 devices Note Refer to ni com support calibrat mancal htm for a copy of the ni671xCal dll file What Is Calibration Calibration consists of verifying the measur...

Страница 2: ...en calibrated with respect to an external standard However external variables such as temperature can still affect measurements The new calibration constants are defined with respect to the calibratio...

Страница 3: ...imize connections and test conditions during calibration Keep connections to the NI 6711 6713 6731 6733 short Long cables and wires act as antennae picking up extra noise which can affect measurements...

Страница 4: ...w to configure the device in NI DAQ Refer to the NI 671X 673X User Manual for detailed installation instructions You can install this manual when you install NI DAQ 1 Power down the computer 2 Install...

Страница 5: ...DAQ User Manual for PC Compatibles at ni com manuals These two documents provide detailed information about using NI DAQ The function reference help includes information on the functions in NI DAQ Th...

Страница 6: ...those specified by the calibration procedure 2 To internally calibrate the device call the Calibrate_E_Series function with the following parameters set as indicated calOP set to ND_SELF_CALIBRATE set...

Страница 7: ...ifying the Performance of the Counter This procedure verifies the performance of the counter The NI 6711 6713 6731 6733 devices have only one timebase to verify so you only need to verify counter 0 Be...

Страница 8: ...nnot modify these values which provides a level of security that ensures users do not accidentally access or modify any calibration constants adjusted by the metrology laboratory This step in the cali...

Страница 9: ...om the device The device is now adjusted with respect to the external source After the device is adjusted you can verify the AO operation by repeating the Verifying Analog Output section Specification...

Страница 10: ...r the test point value should fall between the upper and lower limit values These limits are expressed in volts Counters Because you cannot adjust the resolution of the counter timers these values do...

Страница 11: ...Programs National Instruments NI DAQ NI DAQ Help and the NI DAQ User Manual for PC Compatibles at ni com manuals for additional information on the software structure Verifying Analog Output Figure 1 V...

Страница 12: ...CTR_Control deviceNumber 0 ND_RESET GPCTR_Set_Application deviceNumber 0 ND_PULSE_TRAIN_GNR GPCTR_Change_Parameter deviceNumber 0 ND_COUNT_1 2 GPCTR_Change_Parameter deviceNumber 0 ND_COUNT_2 2 Select...

Страница 13: ...djusting the NI 6711 6713 6731 6733 Figure 3 Adjusting the NI 6711 6713 6731 6733 Calibrate_E_Series deviceNumber ND_SELF_CALIBRATE ND_USER_EEPROM_AREA 0 Get_Cal_Date deviceNumber CalDate Calibrate_E_...

Отзывы: