NI 6711/6713/6731/6733 Calibration Procedure

5.

Call

AO_Configure

to configure the device for the appropriate device

number, channel, and output polarity (the NI 6711/6713/6731/6733
devices support only bipolar output range). Use channel 0 as the
channel to verify. Read the remaining settings from the specification
table for the device.

6.

Call

AO_VWrite

to update the AO channel with the appropriate

voltage. The voltage value is in the specification table.

7.

Compare the resulting value shown by the DMM to the upper and
lower limits on the specification table. If the value falls between these
limits, the device has passed the test.

8.

Repeat steps 3 through 5 until you have tested all values.

9.

Disconnect the DMM from DAC0OUT, and reconnect it to the next
channel, making the connections from Table 1.

10. Repeat steps 3 through 9 until you have verified all channels.

11. Disconnect the DMM from the device.

You have now verified the AO channels of the device.

Verifying the Performance of the Counter

This procedure verifies the performance of the counter. The
NI 6711/6713/6731/6733 devices have only one timebase to verify, so you
only need to verify counter 0. Because you cannot adjust this timebase, you
can only verify the performance of counter 0. Make sure that you have read
the

Equipment and Other Test Requirements

section, and then follow this

procedure:

1.

Connect the counter positive input to GPCTR0_OUT (pin 2) and the
counter negative input to DGND (pin 35).

Note

Pin numbers are given for 68-pin I/O connectors only. If you are using a 50-pin

I/O connector, refer to the device documentation for signal connection locations.

2.

Call

GPCTR_Control

with

action

set to

ND_RESET

to place the

counter in a default state.

3.

Call

GPCTR_Set_Application

with

application

set to

ND_PULSE_TRAIN_GNR

to configure the counter for pulse-train

generation.

4.

Call

GPCTR_Change_Parameter

with

paramID

set to

ND_COUNT_1

and

paramValue

set to

2

to configure the counter to output a pulse

with an off time of 100 ns.

5.

Call

GPCTR_Change_Parameter

with

paramID

set to

ND_COUNT_2

and

paramValue

set to

2

to configure the counter to output a pulse

with an on time of 100 ns.

Содержание NI 6731

Страница 1: ...ed for calibrating NI 6711 6713 6731 6733 devices Note Refer to ni com support calibrat mancal htm for a copy of the ni671xCal dll file What Is Calibration Calibration consists of verifying the measur...

Страница 2: ...en calibrated with respect to an external standard However external variables such as temperature can still affect measurements The new calibration constants are defined with respect to the calibratio...

Страница 3: ...imize connections and test conditions during calibration Keep connections to the NI 6711 6713 6731 6733 short Long cables and wires act as antennae picking up extra noise which can affect measurements...

Страница 4: ...w to configure the device in NI DAQ Refer to the NI 671X 673X User Manual for detailed installation instructions You can install this manual when you install NI DAQ 1 Power down the computer 2 Install...

Страница 5: ...DAQ User Manual for PC Compatibles at ni com manuals These two documents provide detailed information about using NI DAQ The function reference help includes information on the functions in NI DAQ Th...

Страница 6: ...those specified by the calibration procedure 2 To internally calibrate the device call the Calibrate_E_Series function with the following parameters set as indicated calOP set to ND_SELF_CALIBRATE set...

Страница 7: ...ifying the Performance of the Counter This procedure verifies the performance of the counter The NI 6711 6713 6731 6733 devices have only one timebase to verify so you only need to verify counter 0 Be...

Страница 8: ...nnot modify these values which provides a level of security that ensures users do not accidentally access or modify any calibration constants adjusted by the metrology laboratory This step in the cali...

Страница 9: ...om the device The device is now adjusted with respect to the external source After the device is adjusted you can verify the AO operation by repeating the Verifying Analog Output section Specification...

Страница 10: ...r the test point value should fall between the upper and lower limit values These limits are expressed in volts Counters Because you cannot adjust the resolution of the counter timers these values do...

Страница 11: ...Programs National Instruments NI DAQ NI DAQ Help and the NI DAQ User Manual for PC Compatibles at ni com manuals for additional information on the software structure Verifying Analog Output Figure 1 V...

Страница 12: ...CTR_Control deviceNumber 0 ND_RESET GPCTR_Set_Application deviceNumber 0 ND_PULSE_TRAIN_GNR GPCTR_Change_Parameter deviceNumber 0 ND_COUNT_1 2 GPCTR_Change_Parameter deviceNumber 0 ND_COUNT_2 2 Select...

Страница 13: ...djusting the NI 6711 6713 6731 6733 Figure 3 Adjusting the NI 6711 6713 6731 6733 Calibrate_E_Series deviceNumber ND_SELF_CALIBRATE ND_USER_EEPROM_AREA 0 Get_Cal_Date deviceNumber CalDate Calibrate_E_...

Отзывы: