NI 6703/6704 Calibration Procedure for Traditional NI-DAQ
12
ni.com
Flowcharts
The following flowcharts show the appropriate NI-DAQ function calls for
verifying and adjusting the NI 6703/6704. Refer to the
section of this document, the
Traditional NI-DAQ Function
Reference Help,
and the
NI-DAQ User Manual for PC Compatibles
for
additional information about the software structure.
Voltage Output Verification
Figure 1.
Voltage Output Verification
Table 2.
Current Output Specifications (NI 6704 Only)
Test Point
Value (mA)
24-Hour Ranges
1-Year Ranges
Lower Limit
(mA)
Upper Limit
(mA)
Lower Limit
(mA)
Upper Limit
(mA)
0.1
0.0985616
0.1014384
0.09855
0.10145
20
19.997885
20.002115
19.995565
20.004435
Refer to Table 1,
Voltage Output Specifications,
to determine the appropriate ranges for
the voltages and currents to verify.
Get_Cal_Date (deviceNumber)
(optional)
Connect the DMM to voltage channel 0.
AO_VWrite(deviceNumber,0,voltage)
Verify that the voltage read by the DMM is within range.