Chapter 1
Introduction—Document Scope
1-2
ni.com
Why Should You Calibrate?
The characteristics of electronic components drift with time and
temperature, which can affect output accuracy as the device ages.
Calibration restores your device to NI standards.
How Often Should You Calibrate?
The output accuracy requirements of your application determine
the calibration interval of your NI 54
XX
device. See the recommended
calibration schedule in Table 1-1. Refer to Chapter 2,
, at any
time to determine if your device needs a full calibration.
Equipment and Other Test Requirements
This section describes the equipment, software, documentation, and test
conditions required for calibrating your NI 54
XX
device.
Test Equipment
Internal adjustment does not require any test equipment. External
adjustment requires different equipment for each type of measurement.
Table 1-1.
Recommended Calibration Schedule
Calibration Type
Calibration Time
Interval
Calibration
Temperature
Interval
Internal
1 week
±5
°
C
External
1 year
—