Chapter 3
Adjustment
3-2
ni.com
Table 3-1.
Measuring Calibration Output
Constant to Adjust
Type of Measurement
Output Connection
Measurement
Device
A0–A6
DC voltage
ARB OUT/VIDEO OUT
DMM
A7–A9
Magnitude of a sinusoidal
voltage (either amplitude,
peak-to-peak amplitude,
or RMS value)
ARB OUT/VIDEO OUT
DMM
Internal reference
DC voltage
ARB OUT/VIDEO OUT
DMM
Frequency response
Magnitude of a sinusoidal
voltage (either amplitude,
peak-to-peak amplitude,
or RMS value)
ARB OUT/VIDEO OUT
Scope/Power
Meter
Duty cycle
Duty cycle
SYNC OUT
Counter
VCXO
Frequency
SYNC OUT
Counter