NI Digital Multimeters Getting Started Guide
5.
Locate the 5
×
20 mm glass fuse shown in Figure 24.
Figure 24.
Removing the Fuse from the NI PCI-4060
6.
Remove the fuse, and verify the fuse is blown by measuring discontinuity across the fuse using
another meter.
7.
Press the new fuse into the silver holding fixture until you hear a snap.
Caution
Do
not
operate the NI PCI-4060 without both insulators replaced and secured.
8.
Reattach the top and bottom insulators in the opposite order you removed them.
1
Glass Fuse
1