CVI, LabVIEW, National Instruments, NI, ni.com, the National Instruments corporate logo, and the Eagle logo are trademarks of National Instruments
Corporation. Refer to the
Trademark Information
at
ni.com/trademarks
for other National Instruments trademarks. The mark LabWindows is
used under a license from Microsoft Corporation. Windows is a registered trdemark of Microsoft Corporation in the United States and other countries.
Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering
National Instruments products/technology, refer to the appropriate location:
Help»Patents
in your software, the
patents.txt
file on your media,
or the
National Instruments Patent Notice
at
ni.com/patents
. Refer to the
Export Compliance Information
at
ni.com/legal/
export-compliance
for the National Instruments global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other
import/export data.
© 2011–2013 National Instruments. All rights reserved.
371897B-01
Jan13
5.
Use the DAQmx Adjust SCXI Calibration VI and select 1503 Current to store the excitation current
constant.
6.
Use the DAQmx Close External Calibration VI to commit the constant overwritten.
7.
Repeat steps 1 through 6 for all of the other channels.
8.
Once you have completed adjusting the SCXI-1503, reverify the module.
Where to Go for Support
The National Instruments Web site is your complete resource for technical support. At
ni.com/
support
you have access to everything from troubleshooting and application development self-help
resources to email and phone assistance from NI Application Engineers.
National Instruments corporate headquarters is located at 11500 North Mopac Expressway, Austin,
Texas, 78759-3504. National Instruments also has offices located around the world to help address your
support needs. For telephone support in the United States, create your service request at
ni.com/
support
and follow the calling instructions or dial 512 795 8248. For telephone support outside the
United States, visit the Worldwide Offices section of
ni.com/niglobal
to access the branch office
Web sites, which provide up-to-date contact information, support phone numbers, email addresses, and
current events.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxAdjust1503CurrentCal
with the following parameters:
calHandle
:
calHandle
channelName
: Channel under test
measOutput
: Result from DMM in
step 3.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxCloseExtCal
with the
following parameters:
calHandle
:
calHandle
action
:
DAQmx_Val_Action_commit