National Instruments Module SCXI-1503 Скачать руководство пользователя страница 15

 

CVI, LabVIEW, National Instruments, NI, ni.com, the National Instruments corporate logo, and the Eagle logo are trademarks of National Instruments 
Corporation. Refer to the 

Trademark Information 

at 

ni.com/trademarks

 for other National Instruments trademarks. The mark LabWindows is 

used under a license from Microsoft Corporation. Windows is a registered trdemark of Microsoft Corporation in the United States and other countries. 
Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering 
National Instruments products/technology, refer to the appropriate location: 

Help»Patents

 in your software, the 

patents.txt

 file on your media, 

or the 

National Instruments Patent Notice

 at

ni.com/patents

. Refer to the 

Export Compliance Information 

at 

ni.com/legal/

export-compliance

 for the National Instruments global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other 

import/export data.

© 2011–2013 National Instruments. All rights reserved.

371897B-01

Jan13

5.

Use the DAQmx Adjust SCXI Calibration VI and select 1503 Current to store the excitation current 
constant.

6.

Use the DAQmx Close External Calibration VI to commit the constant overwritten.

7.

Repeat steps 1 through for all of the other channels.

8.

Once you have completed adjusting the SCXI-1503, reverify the module.

Where to Go for Support

The National Instruments Web site is your complete resource for technical support. At 

ni.com/

support

 you have access to everything from troubleshooting and application development self-help 

resources to email and phone assistance from NI Application Engineers.

National Instruments corporate headquarters is located at 11500 North Mopac Expressway, Austin, 
Texas, 78759-3504. National Instruments also has offices located around the world to help address your 
support needs. For telephone support in the United States, create your service request at 

ni.com/

support

 and follow the calling instructions or dial 512 795 8248. For telephone support outside the 

United States, visit the Worldwide Offices section of 

ni.com/niglobal

 to access the branch office 

Web sites, which provide up-to-date contact information, support phone numbers, email addresses, and 
current events.

LabVIEW Block Diagram

NI-DAQmx Function Call

Call 

DAQmxAdjust1503CurrentCal

with the following parameters:

calHandle

calHandle

channelName

: Channel under test

measOutput

: Result from DMM in 

step 3.

LabVIEW Block Diagram

NI-DAQmx Function Call

Call 

DAQmxCloseExtCal

 with the 

following parameters:

calHandle

calHandle

action

DAQmx_Val_Action_commit

Содержание Module SCXI-1503

Страница 1: ...og Input 8 Verifying the Excitation Current 11 Adjustment 11 Adjusting the Analog Input 12 Adjusting the Excitation Current 14 Where to Go for Support 15 Software Requirements Calibrating the SCXI 1503 requires installing NI DAQmx 8 1 or later on the calibration system You can download NI DAQmx from ni com downloads NI DAQmx 8 1 or later supports self calibration and external calibration in the La...

Страница 2: ...or the SCXI 1503 SCXI 1306 Terminal Block Installation Guide includes detailed information and published specifications for the SCXI 1306 NI DAQmx Help includes information about creating applications that use NI DAQmx NI DAQmx C Reference Help includes information about the functions in the driver You can find the latest versions of the documentation at ni com manuals Calibration Interval Nationa...

Страница 3: ...ange calibrate the module at that temperature Keep relative humidity below 80 Table 1 Recommended Equipment Equipment Recommended Model Parameter Measured Minimum Requirements Calibrator Fluke 5500A Analog Input If this instrument is unavailable use a calibrator that is accurate to 10 ppm 5 digits DMM PXI 4071 Excitation Current Analog Input If this instrument is unavailable use a multiranging 7 d...

Страница 4: ...ial Setup Install the module and configure it in Measurement Automation Explorer MAX 2 Self Calibration Adjust the self calibration constants of the module 3 Verification Verify the existing operation of the module This step confirms whether the module is operating within the published specifications prior to adjustment 4 Adjustment Adjust the calibration constants of the module The adjustment pro...

Страница 5: ...3 SCXI 1000 Chassis 4 SCXI 1349 Adapter 5 SH68 68 EP Cable for E Series DAQ Device SHC68 68 EPM Cable for M or X Series DAQ Device 6 Test Computer with NI PCI e DAQ Device 7 DMM 8 Calibrator 9 Calibrator to SCXI 1306 Wiring 10 CB 50 to DMM Wiring 11 CB 50 Terminal Accessory 12 NB1 Ribbon Cable 10 0000208 V 10 0000001 V DC DCV ACV OHM DCI ACI FRE Q Loca l Auto Cal Auto Zero Trig Rec all Stat e NPL ...

Страница 6: ...DAQ device refer to Figure 2 to make your connections 1 SCXI 1306 Terminal Block 2 SCXI 1503 Module 3 SCXI 1000 Chassis 4 SCXI 1349 Adapter 5 SHC68 68 EPM Cable 6 PXI Chassis 7 DMM 8 Calibrator 9 Calibrator to SCXI 1306 Wiring 10 CB 50 to DMM Wiring 11 CB 50 Terminal Accessory 12 NB1 Ribbon Cable 10 0000208 V 10 0000001 V DC DCV ACV OHM DCI ACI FRE Q Loca l Auto Cal Auto Zero Trig Rec all Stat e N...

Страница 7: ...1000 chassis 1 Configuration Switches shown in Off position 2 Switch Legends 3 Ground Lug 1 2 1 1 2 1 ON Closed IEX AI IEX AI Ch X IEX AI AI IEX IEX AI AI IEX CH 0 CH 1 CH 2 CH 3 CH 4 CH 5 CH 6 CH 7 CH 9 CH 8 CH 11 CH 10 CH 13 CH 12 CH 15 CH 14 1 2 3 4 5 6 7 8 1 2 3 4 5 6 7 8 CH 4 CH 5 CH 6 CH 7 CH 0 CH 1 CH 2 CH 3 CH 12 CH 13 CH 14 CH 15 CH 8 CH 9 CH 10 CH 11 ON ON SW 1 SW 2 ON ON SW 3 SW 4 1 1 2...

Страница 8: ...ning the module you want to calibrate 6 Select the module that you want to calibrate 7 Initialize a self calibration of the module you want to calibrate Verification This section provides instructions for verifying the following SCXI 1503 specifications Analog input Excitation current Note Limits in Table 2 are based on the March 2007 edition of the SCXI 1503 User Manual which includes the specifi...

Страница 9: ...Q device E Series have two scaling coefficients while M Series and X Series have four scaling coefficients Record the first two scaling coefficients and save the values as a0 and a1 Note The DAQ device scaling coefficients a0 and a1 are the same for each channel on the SCXI module since the DAQ device multiplexes all SCXI channels to channel ai0 on the DAQ device 8 Clear the task 1 To DMM V 2 To D...

Страница 10: ...channel MUX to the specified channel 15 Use the saved device scaling coefficients from the DAQ device a0 and a1 from step 7 and the combined DAQ device and SCXI 1503 module b0 and b1 from step 13 to compute the device scaling coefficients for just the SCXI 1503 c0 and c1 16 Configure the DMM to measure voltage at the CB 50 outputs and take a measurement Table 2 Test Point and Limit SCXI 1503 Gain ...

Страница 11: ... 15 nA or better at 100 A Note The PXI 4071 can measure up to 105 A in the 100 A range 3 Create an AI voltage channel and configure it to use internal excitation by setting the AI Excit Src to internal and then read the excitation current value from the AI Excit Val property node Save this value to DUT IEX 4 Configure the DMM to get current measurement results Save this value to DMM Reading 5 Calc...

Страница 12: ... connector 5 Connect the DMM V to pin 3 and DMM V to pin 4 of the CB 50 terminal accessorys Refer to Figure 5 for pin locations Figure 5 CB 50 Terminal Accessory Wiring Diagram for Connection to DMM Note When MAX configures a module it is assigned a module name Each function call uses this module name to determine the correct module to adjust This document uses Dev1 when referring to the module na...

Страница 13: ... at the CB 50 outputs and take a measurement Table 3 DC Source SCXI 1503 Gain Test Voltage 1 9 1 V 6 1 V 3 1 V 0 V 3 1 V 6 1 V 9 1 V 100 0 091 V 0 061 V 0 031 V 0 V 0 031 V 0 061 V 0 091 V LabVIEW Block Diagram NI DAQmx Function Call Call DAQmxInitExtCal with the following parameters deviceName Dev1 password NI or SCXI default calHandle calHandle LabVIEW Block Diagram NI DAQmx Function Call Call D...

Страница 14: ... in the off position as shown in Figure 3 2 Configure the DMM to do a current measurement with a range of 100 A and an accuracy of 15 nA or better at 100 A Note The PXI 4071 can measure up to 105 A in the 100 A range 3 Take a current measurement using the DMM Save this value to DMM Reading 4 Use the DAQmx Initialize External Calibration VI to create a calibration task LabVIEW Block Diagram NI DAQm...

Страница 15: ...to commit the constant overwritten 7 Repeat steps 1 through 6 for all of the other channels 8 Once you have completed adjusting the SCXI 1503 reverify the module Where to Go for Support The National Instruments Web site is your complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and ph...