SCXI-1503 Calibration Procedure
14
ni.com
10. Use the DAQmx Adjust SCXI Calibration VI to calculate and store the constants.
11. Use the DAQmx Close External Calibration VI to commit the constant overwritten.
12. Repeat steps 6 through 11 for all SCXI-1503 channels.
13. Repeat steps 6 through 12 to test all test points for gain = 1.
14. Repeat steps 7 through 13 using a gain of 100 on the SCXI-1503.
15. Self-calibrate the SCXI-1503.
Adjusting the Excitation Current
Complete the following steps to adjust the excitation current of the SCXI-1503:
1.
Connect IEX
x
+ and IEX
x
- to DMM current input with
x
being the number of the channel from
0 to 15.
Note
Ensure that all switches on the SCXI-1306 are in the off position as shown in Figure 3.
2.
Configure the DMM to do a current measurement with a range of 100
A and an accuracy of 15 nA
or better at 100
A.
Note
The PXI-4071 can measure up to 105
A in the 100
A range.
3.
Take a current measurement using the DMM. Save this value to DMM Reading.
4.
Use the DAQmx Initialize External Calibration VI to create a calibration task.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxAdjust1503Cal
with
the following parameters:
calHandle
:
calHandle
refVoltage
: test point sent out by the
calibrator.
measOutput
: DMM measurement
result in step 9.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxCloseExtCal
with the
following parameters:
calHandle
:
calHandle
action
:
DAQmx_Val_Action_
commit
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxInitExtCal
with the
following parameters:
deviceName
:
Dev1
password
:
NI
or
SCXI
calHandle
:
&calHandle