SCXI-1503 Calibration Procedure
10
ni.com
9.
Configure the calibrator to output a DC voltage. Table 2 shows the test point values.
10. Create an AI voltage channel with an input range of ±10 V for one channel on the SCXI-1503.
11. Configure the SCXI-1503 input channel for a finite acquisition with a sample rate of 1000 Hz and
1000 samples per channel.
12. Use DAQmx properties to set a gain of 1.
13. Use DAQmx properties to read the DAQ device scaling coefficients. E Series have two scaling
coefficients while M Series and X Series have four scaling coefficients. Record the first two scaling
coefficients and save the values as
b
0
and
b
1
.
14. Start the task to set the channel MUX to the specified channel.
15. Use the saved device scaling coefficients from the DAQ device,
a
0
and
a
1
from step 7, and the
combined DAQ device and SCXI-1503 module,
b
0
and
b
1
from step 13, to compute the device
scaling coefficients for just the SCXI-1503,
c
0
and
c
1
.
16. Configure the DMM to measure voltage at the CB-50 outputs and take a measurement.
Table 2.
Test Point and Limit
SCXI-1503 Gain
Test Point Level (V)
Lower Limit (V)
Upper Limit (V)
1
9
8.991680513
9.008319487
6
5.993900513
6.006099487
3
2.996120513
3.003879487
0
-0.001659487
0.001659487
-3
-3.003879487
-2.996120513
-6
-6.006099487
-5.993900513
-9
-9.008319487
-8.991680513
100
0.09
0.089898305
0.090101695
0.06
0.059920505
0.060079495
0.03
0.029942705
0.030057295
0
-0.0000350949
0.0000350949
-0.03
-0.030057295
-0.029942705
-0.06
-0.060079495
-0.059920505
-0.09
-0.090101695
-0.089898305
c
0
b
0
b
1
a
1
-----
–
a
0
=
c
1
b
1
a
1
-----
=