NI 4070/4072 Calibration Procedure
16
ni.com
5.
Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_AC_VOLTS
•
Range
=
0.05
•
Resolution
=
50e–9
6.
Call
niDMM_ConfigureAutoZeroMode w
ith the following
parameters:
•
Auto Zero Mode
=
NIDMM_VAL_AUTO_ZERO_ON
7.
Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 18.
8.
Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_AC_VOLTS_DCCOUPLED
•
Range
=
0.05
•
Resolution
=
50e–9
9.
Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 18.
10. Output 50 mV at 30 Hz on the calibrator.
11. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_AC_VOLTS_DCCOUPLED
•
Range
=
0.05
•
Resolution
=
50e–9
12. Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 18.
13. Refer to Table 3 for the appropriate calibrator outputs and parameter
values as you complete the following steps:
a.
On the calibrator, output the value listed in the
Calibrator Output
column in Table 3 for the current iteration.
b.
Call
niDMM_ConfigureMeasurement
with
Mode
set to
NIDMM_VAL_AC_VOLTS
and the remaining parameters as shown
in Table 3 for the current iteration.
c.
Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 18.
d.
Call
niDMM_ConfigureMeasurement
again, changing
Mode
to
NIDMM_VAL_AC_VOLTS_DCCOUPLED
.
e.
Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 18.