NI cDAQ Chassis Calibration Procedure

2

ni.com

NI cDAQ-9188 Ethernet Chassis Quick Start

—This document explains how to install the 

application and NI-DAQmx driver software, set up your NI cDAQ-9188 Ethernet chassis and 
I/O modules, confirm the system is operating properly, and take your first measurement. You can 
find this document at 

ni.com/manuals

.

Calibration Interval

National Instruments recommends a calibration interval of one year for the NI cDAQ chassis. Adjust the 
recommended calibration interval based on the measurement accuracy demands of your application.

Test Equipment

National Instruments recommends that you use the equipment in Table 1 for calibrating an NI cDAQ 
chassis.

Test Conditions

The following setup and environmental conditions are required to ensure the NI cDAQ chassis meets 
published specifications.

Keep connections to the C Series module as short as possible. Long cables and wires act as 
antennae, picking up extra noise that can affect measurements.

Maintain an ambient temperature of 23 ±5 °C. The device temperature will be greater than the 
ambient temperature.

Keep relative humidity below 90%.

Allow a warm up time of at least 10 minutes to ensure that the measurement circuitry is at a stable 
operating temperature.

Calibration Procedures

The calibration process includes the following steps:

1.

Initial Setup

—Install the device and configure it in Measurement & Automation Explorer (MAX).

2.

Verification

Verify the existing operation of the device. This step confirms whether the device is 

operating within the published specifications prior to adjustment.

3.

Adjustment

—If the device does not fall within the desired specifications, return the device to NI for 

repair. Refer to 

Where to Go for Support

 for assistance in returning the chassis to NI.

4.

Reverification

—Repeat the verification procedure to ensure that the device is operating within the 

published specifications after adjustment.

The first two procedures are described in more detail in the following sections.

Table 1.  

Recommended Equipment

Equipment

Recommended Model

Minimum Requirements

Counter

Agilent 53131A

If this instrument is unavailable, use a 
counter accurate to ±5 ppm.

C Series Digital I/O Module

NI 9401 or NI 9402

C Series Connection Accessory

NI 9934

(NI 9401)

 25-pin DSUB connector kit.

Содержание cDAQ

Страница 1: ...support for the ADE that you intend to use Documentation Requirements For information about NI DAQmx and NI cDAQ chassis you can consult the following documents DAQ Getting Started guides These guides provide instructions for installing and configuring NI DAQ devices NI DAQmx Help This help file contains general information about measurement concepts key NI DAQmx concepts and common applications t...

Страница 2: ...ature will be greater than the ambient temperature Keep relative humidity below 90 Allow a warm up time of at least 10 minutes to ensure that the measurement circuitry is at a stable operating temperature Calibration Procedures The calibration process includes the following steps 1 Initial Setup Install the device and configure it in Measurement Automation Explorer MAX 2 Verification Verify the ex...

Страница 3: ... the C Series module Figure 1 NI 9401 and NI 9402 Pin Assignments 2 Create and configure a counter output channel according to Table 2 3 Start the generation of a square wave 4 Configure the counter to measure frequency and use a 1 MΩ impedance 5 Take a measurement of the square wave 6 Compare the counter reading to the limits in Table 3 Table 2 Counter Output Configuration Frequency Scaled Units ...

Страница 4: ...on development self help resources to email and phone assistance from NI Application Engineers National Instruments corporate headquarters is located at 11500 North Mopac Expressway Austin Texas 78759 3504 National Instruments also has offices located around the world to help address your support needs For telephone support in the United States create your service request at ni com support and fol...

Отзывы: