BNC-2110 Installation Guide

6

ni.com

Measuring Floating Signals

You can use the BNC-2110 to measure floating and ground-referenced 
analog input signals. To measure floating signal sources, move the switch 
located below the BNC connector for the AI channel you are using to the 
floating source switch position (labeled FS). In the floating source switch 
position, the amplifier negative terminal connects to ground through a 
5 k

resistor in parallel with a 0.1 

µ

F capacitor. Table 3 shows the 

BNC-2110 switch configuration options.

Measuring Ground-Referenced Signals

To measure ground-referenced signals, move the switch to either 
the floating or ground-referenced source position. For best results, use the 
ground-referenced source position (labeled GS) to avoid ground loops. 
Refer to your DAQ device documentation for more information on 
measuring floating and ground-referenced signals.

Table 3.  

Configuration Summary

BNC-2110 Switch 

Configuration

Signal Source Types for E Series Devices

Floating Source

Ground-Referenced Source

Floating Source 
(FS)

Ground-
Referenced Source 
(GS)

+

Signal

Source

BNC-2110

DAQ Device

(Differential Input

Mode)

AI GND

+

Recommended

+

Signal

Source

BNC-2110

DAQ Device

(Differential Input

Mode)

AI GND

+

Not Recommended

+

Signal

Source

BNC-2110

Improper Configuration

DAQ Device

(Differential Input

Mode)

+

AI GND

AI GND

AI GND

+

Signal

Source

BNC-2110

DAQ Device

(Differential Input

Mode)

AI GND

+

Recommended

Содержание BNC-2110

Страница 1: ...rminal block with 30 pins for signal connections and a 68 pin I O connector that connects to the E Series S Series and waveform generation Multifunction DAQ devices The BNC 2110 is ideal for simplifyi...

Страница 2: ...onnecting the BNC 2110 to Your DAQ Device 1 Place the BNC 2110 near the host computer or use the optional DIN rail mounting kit which you can order from National Instruments For more information about...

Страница 3: ...r off any external signals connected to the BNC 2110 before you power off your computer Caution The BNC 2110 is not designed for input voltages greater than 42 V even if a user installed voltage divid...

Страница 4: ...lences table in the E Series Help at ni com manuals Figure 2 BNC 2110 Front Panel 1 Floating Source FS Ground Referenced Source GS Switch 2 BNC Connectors 3 Terminal Block Retaining Screws 4 Terminal...

Страница 5: ...nnectors you use depends on your DAQ device and your application Table 2 BNC 2110 Connector Signal Descriptions Front Panel BNC Connectors Signal Description AI 0 1 Differential Analog Input channels...

Страница 6: ...the switch to either the floating or ground referenced source position For best results use the ground referenced source position labeled GS to avoid ground loops Refer to your DAQ device documentatio...

Страница 7: ...on the BNC 2110 front panel These terminals and their associated BNC connectors give you the flexibility to choose up to two additional digital timing signals For example if an application requires a...

Страница 8: ...neral Purpose Counters 0 and 1 respectively CTR 1 OUT Counter 1 Output This is the output from General Purpose Counter 1 USER 1 User defined 1 Connected to USER 1 BNC this terminal provides a user def...

Страница 9: ...Part 15 Class A Compliant Note For EMC compliance you must operate this device with shielded cabling CE Compliance This product meets the essential requirements of applicable European Directives as am...

Страница 10: ...mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products refer to the appropriate location Help Patents in your software the pat...

Отзывы: