2. Connect the NI 9224 to the calibrator as shown in the following figure.

Figure 1.

 Accuracy Verification Connections

NI 9224

AI0–

+

Calibrator

AI0+

AI1–

AI1+

AI7–

AI7+

Note

 

The chassis ground lug connection to ground is the NI 9224 ground

connection.

3. Set the calibrator voltage to a Test Point value indicated in the 

NI 9224 Verification Test

Limits

 table.

4. Set the calibrator to Operate mode (OPR).
5. Acquire and average samples for each channel.

a) Create and configure an AI voltage task according to the following 

NI 9224

Accuracy Verification Configuration

 table.

Table 2.

 NI 9224 Accuracy Verification Configuration

Minimum (V) Maximum (V) Scaled Units

Physical Channels

Channel

Configuration

 

 

 

 

 

-10

10

Volts

cDAQ1Mod8/AI0:AI7 Differential

b) Configure the task timing according to the following 

NI 9224 Timing Configuration

table.

Table 3.

 NI 9224 Timing Configuration

Sample Mode

Samples Per Channel

Rate (kS/s)

 

 

 

Finite Samples

100

1

c) Start the task.
d) Read the samples from the NI 9224.
e) Average the samples for each AI channel and record the average.
f) Stop and clear the task.

6. Set the calibrator to Standby mode (STBY).
7. Compare the averaged samples to the limits in the 

NI 9224 Verification Test Limits

 table

to verify the input accuracy. If the value is within these limits, the device passes the test.

8. Repeat steps 3 through 7 for each test point in the 

NI 9224 Verification Test Limits

 table.

9. Disconnect the calibrator from the device.

4

  | 

ni.com

  | 

NI 9224 Calibration Procedure

Содержание 9224

Страница 1: ...NI 9224...

Страница 2: ...ication 6 Accuracy under Calibration Conditions 6 Worldwide Support and Services 6 Software Calibrating the NI 9224 requires the installation of NI DAQmx 18 1 or later on the calibration system You ca...

Страница 3: ...NI DAQmx C properties NI DAQmx NET Help Support for Visual Studio Reference information for NI DAQmx NET methods and NI DAQmx NET properties key concepts and a C enum to NET enum mapping table Test E...

Страница 4: ...ing lug to a 14 AWG 1 6 mm wire b Connect the ring lug to the ground terminal on the side of the NI cDAQ 9178 using the ground screw c Attach the other end of the wire to the system safety ground 4 In...

Страница 5: ...aled Units Physical Channels Channel Configuration 10 10 Volts cDAQ1Mod8 AI0 AI7 Differential b Configure the task timing according to the following NI 9224 Timing Configuration table Table 3 NI 9224...

Страница 6: ...ction 3 Adjust the NI 9224 a Initialize a calibration session on the NI 9224 The default password is NI b Input the external temperature in degrees Celsius c Call the NI 9224 get C Series adjustment p...

Страница 7: ...in the following table are based on calibrated scaling coefficients which are stored in the onboard EEPROM The following accuracy table is valid for calibration under the following conditions Ambient...

Страница 8: ...or your product by visiting ni com certification If your product supports calibration you can obtain the calibration certificate for your product at ni com calibration NI corporate headquarters is loc...

Страница 9: ...s EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and...

Отзывы: