Table 4.

 NI 9224 Verification Test Limits

Test Points

Device Under Test Settings/Configuration

Location

Value

Lower Limit (VDC)

Upper Limit (VDC)

 

 

 

 

Positive full-scale

9.5

9.485

9.515

Mid

0

-0.001161

0.001161

Negative full-scale

-9.5

-9.515

-9.485

Note

 

The test limits listed in the 

NI 9224 Verification Test Limits

 table are derived

using values from the 

NI 9224 Accuracy Under Calibration Conditions

 table.

Adjustment

The following performance adjustment procedure describes the sequence of operation required

to adjust the NI 9224.

Accuracy Adjustment

Complete the following procedure to adjust the accuracy performance of the NI 9224.
1. Set the calibrator to Standby mode (STBY).
2. Connect the NI 9224 to the calibrator as shown in the 

Accuracy Verification

 section.

3. Adjust the NI 9224.

a) Initialize a calibration session on the NI 9224. The default password is 

NI

.

b) Input the external temperature in degrees Celsius.
c) Call the NI 9224 get C Series adjustment points function to obtain an array of

recommended calibration voltages for the NI 9224.

d) Set the calibrator to a reference value determined by the array of recommended

calibration voltages.

e) Set the calibrator to Operate mode (OPR).
f) Call and configure the NI 9224 adjustment function according to 

NI 9224

Adjustment Configuration

 table.

Table 5.

 NI 9224 Adjustment Configuration

Physical Channel

Reference Value

 

 

cDAQ1Mod8/AI0:7

A reference value from the array of adjustment points

g) Set the calibrator to Standby mode (STBY).
h) Repeat steps d through g for each point in the array.
i)

Close the calibration session.

j)

Close and commit the calibration session.

4. Disconnect the NI 9224 from the calibrator.

NI 9224 Calibration Procedure 

  | 

© National Instruments

  | 

5

Содержание 9224

Страница 1: ...NI 9224...

Страница 2: ...ication 6 Accuracy under Calibration Conditions 6 Worldwide Support and Services 6 Software Calibrating the NI 9224 requires the installation of NI DAQmx 18 1 or later on the calibration system You ca...

Страница 3: ...NI DAQmx C properties NI DAQmx NET Help Support for Visual Studio Reference information for NI DAQmx NET methods and NI DAQmx NET properties key concepts and a C enum to NET enum mapping table Test E...

Страница 4: ...ing lug to a 14 AWG 1 6 mm wire b Connect the ring lug to the ground terminal on the side of the NI cDAQ 9178 using the ground screw c Attach the other end of the wire to the system safety ground 4 In...

Страница 5: ...aled Units Physical Channels Channel Configuration 10 10 Volts cDAQ1Mod8 AI0 AI7 Differential b Configure the task timing according to the following NI 9224 Timing Configuration table Table 3 NI 9224...

Страница 6: ...ction 3 Adjust the NI 9224 a Initialize a calibration session on the NI 9224 The default password is NI b Input the external temperature in degrees Celsius c Call the NI 9224 get C Series adjustment p...

Страница 7: ...in the following table are based on calibrated scaling coefficients which are stored in the onboard EEPROM The following accuracy table is valid for calibration under the following conditions Ambient...

Страница 8: ...or your product by visiting ni com certification If your product supports calibration you can obtain the calibration certificate for your product at ni com calibration NI corporate headquarters is loc...

Страница 9: ...s EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and...

Отзывы: