REMOVING THE ARM SKIN AND VEINS

A.

To remove the skin from the arm core, open the shoestring at the top of the arm.

B.

Remove the inlet and outlet tubes from the arm skin.

C.

Coat the exterior skin lightly with the lubricant and strip off the skin so that it will be inside out when 

you finish. 

(See figure 1.)

 

Note: 

If you are replacing the skin and the veins, simply cut the skin free from the arm core using 

scissors and discard. If replacing only the veins, pull the skin down only far enough to expose the 

veins without removing the skin completely from the fingers. This will make it easier to pull the skin 

back on when finished.

D.

Remove the foam shoulder injection pad. 

(See figure 2.)

E. 

Remove the bicep muscle located over the manifold.

F. 

Remove the old “veins” from the grooves, unplug them from the manifold, and discard them.

G.

If needed, clean the arm core with soap and water.

Figure 1

Figure 2

B. 

 Plug one end of tubing into fitting #1. 

(See figure 4 – 

Red

.)

 

Note:

 A small amount of lubricant or water will make it easier to slide the tubing over the connector. 

Also, the plastic wrap around the tubing may need to be pulled back from the end of the tubing 1/2". Work 

the tubing completely into the channel over the back of the arm, around the third finger from the inside, 

and back over the arm in the same channel. Connect the end of the tubing into fitting #4. Veins may be 

trimmed with scissors if necessary for a good fit; just make sure that the connector block is back in its 

correct location within the core and that the tubing is not kinked, pinched, or too tightly stretched. Slight 

pinching in the finger area is normal and will not obstruct blood flow.

REPLACING THE VEIN TUBING

A.

Figure 3 shows how the four vein tubing sections should 

be connected to the connector block. The tube fittings are 

not shown.

Note: 

All diagrams have been color coded for illustration 

purposes only. The latex veins you received with your kit are 

all the same color and length and may be used inter-

changeably.

Figure 3

Figure 4

G.

Feed the inlet/outlet tubing through the skin to the outside.

H. 

 Adjust the skin over the upper shoulder and retie the lace to hold it in position. Rinse the lubricant 

from the skin with warm water. 
If the skin stretches slightly during use, it can be tightened by retying the lace.

SUPPLIES AND REPLACEMENT PARTS

LF00845U

 

Life

/

form

®

 Blood, 1 quart

LF00846U

 

Life

/

form

®

 Blood, 1 gallon

LF00991U

  Shoulder Injection Pad Replacement

LF00993U

  Replacement Vein Manifold

LF00985U

  Lubricant, pack of 6, 2-oz. bottles

LF00996U

  Vyna-Bond Adhesive

LF01022U

  Fluid Supply Stand

LF01099U

  Vein Tubing Sealant

LF01104U

 

Life

/

form

®

 IV Arm Skin & Vein Replacement DVD — FREE

LF01105U

 

Life

/

form

®

 IV Arm Set-Up & Troubleshooting DVD — FREE

LF01122U

 

Light Skin & Vein Replacement Kit

LF01123U

 

Dark Skin & Vein Replacement Kit

LF01124U

  Vein Replacement Kit

LF01130U

  Fluid Supply Bag

LF09919U

  Nasco Cleaner

Figure 1

Figure 2

Actual product may vary slightly from photo. Nasco reserves the right to change product color, materials, 

supplies, or function as needed.

HC18000028INM39_OldMani.indd   5-6

11/15/18   9:06 AM

Содержание LF01122U

Страница 1: ...theterization TPN LF01036U Spinal Injection LF01037U Hemodialysis Practice Arm LF01087U Central Venous Cannulation LF01108U Infant Intraosseous Infusion LF01121U Advanced IV Arm LF01131U Venipuncture and Injection Arm LF01139U Advanced IV Hand LF01162U Venatech IV Trainer LF01184U Venatech IM Sub Q LF03616U Complete Child CRiSis Manikin LF03617U Deluxe Child CRiSis Manikin with Arrhythmia Tutor LF...

Страница 2: ...r IV arm has black tubing connectors discard the eight tubing adapters and proceed with just the replacement tubing If the manifold on your IV arm has two black and eight white tubing connectors you will need to use the new adapters included in the kit To connect the new adapters simply slide the eight adapters over the connec tors after the old veins have been removed After you are finished conne...

Страница 3: ... the connector block is back in its correct location within the core and that the tubing is not kinked pinched or too tightly stretched Slight pinching in the finger area is normal and will not obstruct blood flow REPLACING THE VEIN TUBING A Figure 3 shows how the four vein tubing sections should be connected to the connector block The tube fittings are not shown Note All diagrams have been color ...

Страница 4: ...large channel and out the hole in the upper arm Replace the muscle pad over it B Pour the remaining lubricant into the skin allowing lubricant to run down the sides See figure 7 Squeeze the skin between extended hands Work your hands back and forth up and down the skin to coat the inside of the skin thoroughly C Grasping the skin with both hands slide the skin down over the core until the fingers ...

Страница 5: ...large channel and out the hole in the upper arm Replace the muscle pad over it B Pour the remaining lubricant into the skin allowing lubricant to run down the sides See figure 7 Squeeze the skin between extended hands Work your hands back and forth up and down the skin to coat the inside of the skin thoroughly C Grasping the skin with both hands slide the skin down over the core until the fingers ...

Страница 6: ... the connector block is back in its correct location within the core and that the tubing is not kinked pinched or too tightly stretched Slight pinching in the finger area is normal and will not obstruct blood flow REPLACING THE VEIN TUBING A Figure 3 shows how the four vein tubing sections should be connected to the connector block The tube fittings are not shown Note All diagrams have been color ...

Страница 7: ...theterization TPN LF01036U Spinal Injection LF01037U Hemodialysis Practice Arm LF01087U Central Venous Cannulation LF01108U Infant Intraosseous Infusion LF01121U Advanced IV Arm LF01131U Venipuncture and Injection Arm LF01139U Advanced IV Hand LF01162U Venatech IV Trainer LF01184U Venatech IM Sub Q LF03616U Complete Child CRiSis Manikin LF03617U Deluxe Child CRiSis Manikin with Arrhythmia Tutor LF...

Отзывы: