![Nasco Healthcare LF01122U Скачать руководство пользователя страница 5](http://html.mh-extra.com/html/nasco-healthcare/lf01122u/lf01122u_instruction-manual_707824005.webp)
4
5
C.
Refer to Figure 4 (
Green
). Plug one end of the tubing into fitting #2. Press the tubing into the channel,
around the loop near the wrist, and back to the connector block at fitting #3.
D.
Refer to Figure 5 (
Yellow
). Plug one end of the tubing into fitting #7. Work it into the channel down
the forearm, around the palm side of the thumb, and back toward the connector block in the same
channel. Pass the tubing over the first section of “green” tubing and under the second section (refer
to diagram above). Connect the tubing to fitting #5.
Figure 4
Figure 5
E.
Refer to Figure 6 (
Blue
). Plug one end of the tubing into fitting #6. Tuck it under the “
green
” tubing,
then continue to press it into the channel, taking extra care to minimize pinching in the loops around the
two fingers and the back of the hand. Run the tubing under the “
red
” tubing and over both lengths of
“
yellow
” tubing before connecting to fitting #8.
F.
Double check that all the tubing is in the correct channel and free of kinks or other obstructions. Check
that all fitting connections are as tight as possible to prevent leaks.
REASSEMBLING THE ARM
A.
Feed the inlet/outlet tube through the large channel and out the hole in the upper arm. Replace the
muscle pad over it.
B.
Pour the remaining lubricant into the skin, allowing lubricant to run down the sides
(See figure 7.)
Squeeze the skin between extended hands. Work your hands back and forth, up and down the skin
to coat the inside of the skin thoroughly.
C.
Grasping the skin with both hands, slide the skin down over the core until the fingers are approach-
ing the finger holes
(See figure 8.)
(Caution: work the skin on gently — pulling hard on the end of
the skin can stretch or tear the material.) As the skin is worked down over the core, be sure that the
tubing remains in the channels.
D.
Guide the fingers and thumb carefully into the proper holes. Mistakes can be corrected by sliding
the skin back an inch or two from the fingers and beginning again. If air has been trapped in the
fingertips, make a small vent hole beneath the fingernail with a 22-gauge needle.
E.
After the fingers are positioned in the hand, slide the skin over the remainder of the core to provide a
snug fit.
(See figure 9.)
Figure 6
F.
Feel along the tubing channel to make sure that the tubing is correctly seated. Tubing that has
moved can usually be moved back into position by working on the outside of the skin.
(See figure 10.)
Figure 7
Figure 8
Figure 9
Figure 10
HC18000028INM39_OldMani.indd 7-8
11/15/18 9:06 AM