Setup and Panel Instructions
3-11
3.8.1 External
attenuator
In case an external attenuator is used to protect the input of the analyzer,
the value of this attenuator can be entered, in order to have the
PMM 9010F taking into account the value in all calculations of measured
levels.
The way to enter the value of the attenuator is the same adopted in all the
other modes (see for examples §4.2 of PMM 9010F User’s Manual).
The inserted value will be displayed together with the others automatically
adjusted by the receiver; all such attenuations values will be resumed
between brackets at the top right of the screen, where the External
Attenuator value will appear as the last shown number.
The 1
st
and the 2
nd
values between brackets are automatically assigned by
the system, according to the selected Limit, in order to guarantee the
proper input sensitivity and dynamic defined by the Standard.
The 1
st
value refers to the attenuator located into the 9010F receiver, while
the 2
nd
value is the attenuator located into the CA0010.
The 3
rd
value refers to the external attenuator.
The external attenuator parameter can be used also to let the system
take into account of cables and other ancillaries cumulative losses.
Содержание PMM CA0010
Страница 8: ...VIII Safety considerations This page has been left blank intentionally...
Страница 18: ...1 10 General Information This page has been left blank intentionally...
Страница 24: ...2 6 Installation This page has been left blank intentionally...
Страница 49: ...Sweep Mode Operating Instructions 4 3 Fig 4 4 Installation folder Fig 4 5 Ready to install...
Страница 72: ...6 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally...
Страница 74: ...Suggerimenti Commenti Note Suggestions Comments Note...