Operating Instruction 

3-7 

 

 

3.2.4 Manual mode 
 
 

 
 
 
 
 
 

 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Manual mode 
stand-by 
 

Using  this  function  it  is  possible  to  do  a  test  and  change  the  field  level 
during  test,  this  feature  can  help  to  measure  the  level  of  susceptibility  of 
the equipment under test. 
Pressing the MAN key the display will show the following: 

Choosing  10A  key  or  100A  key  the  PMM  1008  will  enable  the  output 
current,  the  operate  led  will  switch  on  and  the  display  will  show  the 
following: 
In  the  second  row  left  side  there  is  the  time  increasing  indication,  in  the 
right side the indication of the field. 

Pressing the [+] and [-] keys it is possible to increase or decrease the field 
level. If the band selected is 100 A the maximum current is 160 A and the 
maximum field depends from the antenna factor.  
If  the  band selected is 10 A the maximum current is 16 A, the maximum 
field depends from the antenna factor but the resolution using the [+] and [-
] key is higher. 
 
Pressing the STBY key the PMM 1008 will disable the output current, the 
operate led will switch off and will show the following display: 
The [+] and [-] keys are disabled, pressing the ON key the PMM 1008 will 
go  back  in  the  previous  menu  and  pressing  the  BACK  key  will  stop  the 
manual mode and will go back in main menu. 

Содержание PMM 1008

Страница 1: ...ou can find the Series Number on the rear panel of the instrument Series Number is in the form 0000X00000 The first four digits and the letter are the Series Number prefix the last five digits are the...

Страница 2: ...vice Nevertheless even this product will eventually become obsolete When that time comes please remember that electronic equipment must be disposed of in accordance with local regulations This product...

Страница 3: ...llation 2 10 Safety Interlock Page 2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 3 2 3 2 8 3 Operation instructions 3 1 Starting PMM 1008 3 2 Main menu display 3 2 1 Continuous test mode 3 2 2 Short test mode 3 2 3 E...

Страница 4: ...1008 with Circular loop 40cm Characteristics of the field generated by a square induction coil 3dB area of the field generated by a square induction coil 3dB area of the field generated by a square in...

Страница 5: ...to it In case of connection of the device to the power mains make sure before connection that the voltage selected on the voltage switch and the fuses are adequate for the voltage of the actual mains...

Страница 6: ...035 Cisano sul Neva SV ITALY based on the following harmonized European Standards successfully applied Safety EN 61010 1 undated reference applies to all editions EMC EN 61326 1 undated reference appl...

Страница 7: ...onsidering the antenna factor the system includes the antenna formed by one coil and the antenna mast The correct output current is automatically set and regulated by the generator PMM 1008 is designe...

Страница 8: ...rrent Output voltage Without load Output current Max continuos Max short Modes Continuous Short Manual Ramp Internal timer Continuous mode Short mode Timer resolution Continuous mode Short mode Size G...

Страница 9: ...General Information 1 3 Fig 1 1 PMM 1008...

Страница 10: ...1 4 General Information 1 7 Front panel Fig 1 2 Front Panel Legend 1 LCD display 2 Functional keys 3 Operate led 4 Stand by led 1 2 3 4...

Страница 11: ...General Information 1 5 1 8 Rear panel Fig 1 3 Rear Panel Legend 1 Mains plug IEC CEE22 2 power switch 3 RS232 485 interface DB9 4 User I O port DB15 5 current output connector 14 mm bolt 5 3 1 2 4...

Страница 12: ...1 6 General Information This page has been left blank intentionally...

Страница 13: ...referring to the accessories check list enclosed with the Users Manual Notify any damage to the carrier personnel as well as the NARDA Representative This is a Safety Class I equipment it is provided...

Страница 14: ...d to NARDA for service please complete the service questionnaire enclosed with the Users Manual and attach it to the instrument To minimize the repair time be as specific as possible when describing t...

Страница 15: ...ne acid acetone or similar matter to avoid damage to external enclosure 2 9 System Installation PMM 1008 is a system with different parts and an installation needing To install the PMM 1008 following...

Страница 16: ...2 4 Installation 3 Lock the loop to the support using the six screws...

Страница 17: ...Installation 2 5 4 Lock the antenna mast on the Cart using the screw on the bottom...

Страница 18: ...ition desired using the screw mounted on the support The final mounting should be the following 6 Put the generator on his support on the Cart 7 Connect the cables from the loop to the output current...

Страница 19: ...antenna produces magnetic field the field is high in the center of the loop set the generator positioning outside of the loop like in the previous figure Fig 2 1 PMM 1008 with Square loop 1m Fig 2 2 P...

Страница 20: ...ock WARNING The PMM 1008 is provided by a safety interlock connected at user port on the rear panel The connections are the following To use the PMM 1008 is necessary to insert the HD Key The interloc...

Страница 21: ...the PMM 1008 For connect the instrument at distribution network don t use extension lead or adapter Connect the power supply cable and HD Key to the rear panel connector and switch on the standby led...

Страница 22: ...ear in the top of the display Select your field and press ON to start the test Pressing the SET key the display will show the following Setting LEVX open level Using this menu it is possible to set th...

Страница 23: ...l show the following In the second row at the left there is the indication of time in seconds In the second row at the right there is the indication of the field calculated considering the output curr...

Страница 24: ...is the indication of the level with reference to the standard The levels available are Level 4 300 A m Level 5 1000 A m if antenna factor is 0 87 otherwise is 1000 0 87 K Level X open in the range 300...

Страница 25: ...tor variac to set the correct current this position will be stored for the test Pressing the BACK key will return in previous menu setting open level and test time From this menu it is possible to set...

Страница 26: ...there is the indication of the decreasing time in seconds Starting the test after pressed the ON key the PMM 1008 will move the internal variac at the position stored during the calibration After thi...

Страница 27: ...tion in the right side the indication of the field Pressing the and keys it is possible to increase or decrease the field level If the band selected is 100 A the maximum current is 160 A and the maxim...

Страница 28: ...stored position and then will regulate around it The calibration can be done for the continuous test CONT or for the short test SHRT or both Pressing CONT key the PMM 1008 will calibrate all the cont...

Страница 29: ...CT key from the setup menu the display will be the following Pressing and keys the PMM 1008 will change the antenna factor When it is changed an antenna factor it is necessary to do a calibration beca...

Страница 30: ...Instruction 3 2 7 Error message If the antenna is disconnected after some seconds the PMM 1008 will display the error message LOOP OPEN In this case check all the connections between the generator an...

Страница 31: ...2 Short duration LEVEL MAGNETIC FIELD LEVEL MAGNETIC FIELD 1 1 1 Not applicable 2 3 2 Not applicable 3 10 3 Not applicable 4 30 4 300 5 100 5 1000 4 2 Induction coil The PMM 1008 is supplied with a 1...

Страница 32: ...d by a square induction coil 1m side in its plane Fig 4 3 3dB area of the field generated by a square induction coil 1m side in the mean orthogonal plane component orthogonal to the plane of the coil...

Страница 33: ...y Communication is done at 9600 Baud with 8 bit 1 stop bit and no parity The communication uses strings with variable byte width The characters used inside the strings are in ASCII format 00 127 at 7...

Страница 34: ...ield level Note that this command is not recognized if received within sub menus 00C2 3 A m will be generated 00Sn where n is the level in the range 4 5 which refer to the related field X is the open...

Страница 35: ...ter was last read Note that once this register is read it is automatically reset The result has the following meaning 1 The argument supplied is out of range or inadequate 4 Menu ERROR A command has n...

Страница 36: ...5 4 How to program This page has been left blank intentionally...

Страница 37: ...ents For the software to operate properly the user s Personal Computer should be equipped with the following 486 Processor or Pentium 16 Mb of RAM at least 10 Mb of free space on hard disk 1 free seri...

Страница 38: ...he PMM 1008 5 When the upgrade procedure is completed correctly the software will display the following message PMM 1008 is now updated with the new version of the internal firmware It is now possible...

Страница 39: ...like to submit to the attention of our service organization is kept in great consideration Moreover we are continuously improving our quality but we know this is a never ending process We would be gl...

Страница 40: ...Suggerimenti Commenti Note Suggestions Comments Note...

Отзывы: