Page 10
3
Safety instructions
•
Read this entire manual of operating instructions carefully before putting the
device into operation. Pay special attention to the marked instructions.
•
This equipment is intended for use by healthcare professionals only.
•
The automatic assessment is not allowed to be the sole criterion of a narcosis
or sedation controlling function. With respect to the dosing of narcotics and
sedatives, the overall clinical findings must be taken into account.
•
Prior to each application, check the functional safety and proper condition of
the device. A damaged device must not be put into operation.
•
For safe use, connect the Narcotrend-Compact M - power supply to a
protective earth conductor.
•
Use only the power cords supplied by MT MonitorTechnik GmbH & Co. KG.
•
For putting into operation use only the components that belong to the device.
Only original parts from MT MonitorTechnik GmbH & Co. KG must be used in
the system.
•
The Narcotrend-Compact M - power supply is an essential part of the device
Narcotrend-Compact M. No other power supply must be used.
•
All the consumables that are used (e.g. electrodes, preparation paste, etc.)
must comply with the equivalent safety requirements and/or harmonized
national standards.
•
The use of a defibrillator is restricted. The electrodes must not be placed
between the defibrillator pads.
•
Before putting the device into operation, check the connection cables for any
signs of damage. Damaged cables and plug-and-socket devices must be
replaced.
•
Operating the device beyond the operation temperature range is not
permissible (see technical documentation). Do not cover the air vents.
•
No liquids must be penetrating into the device. Whenever any liquid does
penetrate into the device, the device must not be put into operation again until
it has been properly checked by an expert.
•
The device is not intended to be used in the presence of oxygen enriched
environments or flammable anesthetic mixtures.
•
Whenever the device has been subjected to increased mechanical stress
(dropping, strong vibrations or severe impact), the device must not be put into
operation again until it has been properly checked by an expert.
•
This device may cause radio interference or may disrupt the operation of
nearby equipment. It may be necessary to take mitigation measures such as
reorienting or relocating the equipment or shielding the location.
•
The equipment needs special precautions regarding electromagnetic
compatibility and needs to be installed and put into service according to
information provided in accompanying documents.