23
7.7
Issue or Problem
An empty battery icon is displayed on screen
and an audio alert is played.
Solution
The actuator is not connected to the driver or
damaged. Reconnect the actuator to driver.
If the problem remains, replace with new
actuator (See 7.8).
Solution
The battery requires recharging (See 7.1).
Issue or Problem
The screen displays actuator disconnection
status icon, and an audio alert is played.
TROUBLESHOOTING
Issue or Problem
The device does not turn on.
Solution
Charge the driver using the supplied
charger for 2 hours. If issue is not resolved,
contact the NanoVibronix local distributor.
Issue or Problem
The actuator is not staying secured to the
skin during treatment.
Solution
Turn off the device. Remove the actuator
patch (See 7.7). Replace with a new adhesive
patch (See 7.3).
Operation
7