22
REPLACING THE ACTUATOR
It is recommended to replace the actuator
after 30 treatments.
New actuators can be ordered from
NanoVibronix or your local distributor.
1
Switch OFF the PainShield
®
driver to
prevent the driver from sounding an alert
when it is disconnected from the actuator.
2
Hold the driver in one hand and the
actuator cable plug in the other hand and
pull apart.
(fig.6)
3cm
5-10cm
1
3
4
5
2
3cm
5-10cm
1
3
4
5
2
fig.6
fig.7
3
Connect the new actuator cable plug
into the driver socket, making sure to
orient the plug correctly in relation to the
socket, and that the plug is fully inserted
and secured. (
fig.7)
7.8
7