Figure B-1 -
System grounding for burial in uncased hole (direct burial)
B.3 Grounding the Sensor in a Metal-cased Hole Installation
Use the guidelines in this section to ground a Trillium Compact Posthole seismometer that is
installed in a metal-cased hole.
Lightning surge failures can occur at improperly grounded metal-cased hole installation sites
when the digitizer’s earth ground at the surface becomes charged to a high voltage while the
seismometer is connected to a more neutral voltage at depth.
To minimize this voltage difference, connect the digitizer chassis ground to the borehole casing
at the wellhead. A steel borehole casing provides an excellent earth ground for all equipment
at the station, as well as a ground shield around the seismometer itself. Surge currents will
travel down the outside surface of a steel casing but will not penetrate significantly to the
inside, due to the small skin depth for fast surge transients. See
A non-conductive plastic casing is another option, although this option is usually not preferred
due to poorer mechanical properties (thermal expansion and creep) which may cause low-
Appendix B - Grounding guidelines
16889R10 • 2020-11-12
Page 91
Содержание Trillium Compact
Страница 1: ...User Guide November 12 2020 Strategic Intelligence Fueled by Science...
Страница 76: ...Figure 10 2 Noise floor figure for 120 s models Chapter 10 Performance 16889R10 2020 11 12 Page 76...
Страница 78: ...Figure 10 4 Trillium Compact seismometer performance Chapter 10 Performance 16889R10 2020 11 12 Page 78...
Страница 83: ...Figure 12 2 Bottom view Figure 12 3 Side view Chapter 12 Features and Dimensions 16889R10 2020 11 12 Page 83...
Страница 85: ...Figure 12 6 Side view Chapter 12 Features and Dimensions 16889R10 2020 11 12 Page 85...
Страница 95: ...Figure C 1 Cut away illustration of a packed carrying case Appendix C Carrying Case 16889R10 2020 11 12 Page 95...