Digitizer Input Range
(Vpp)
Digitizer
Sensitivity
(count /µV)
Trillium Compact
Sensitivity
(V/(m/s))
Clip Level
(mm/s)
40
0.4
754
26.5
20
0.8
754
13.3
10
1.6
754
6.63
4
4
754
2.65
2
8
754
1.33
1
16
754
0.66
0.5
32
754
0.33
Table 6-1 -
Clip Level versus Gain Range
Digitizer Input
Range (Vpp)
Centaur System Sensitivity
(Counts/(m/s))
Pegasus System Sensitivity
(Counts/(m/s))
40
2.99e+08
2.99e+08
20
5.99e+08
6.03e+08
10
1.20e+09
1.21e+09
4
2.99e+09
3.02e+09
2
5.99e+09
6.03e+09
1
1.20e+10
1.21e+10
0.5
2.40e+10
2.41e+10
Table 6-2 -
System Sensitivity versus Gain Range
Note that the exact nominal system sensitivity will vary depending on the normalization
frequency (typically 1 Hz) and the lower corner period. The exact nominal sensor sensitivity for
a particular type of Trillium Compact seismometer can be taken from the appropriate table in
"Frequency Response for 120 s and 20 s models" on page 72
Chapter 6 - Configuring Your System
16889R10 • 2020-11-12
Page 59
Содержание Trillium Compact
Страница 1: ...User Guide November 12 2020 Strategic Intelligence Fueled by Science...
Страница 76: ...Figure 10 2 Noise floor figure for 120 s models Chapter 10 Performance 16889R10 2020 11 12 Page 76...
Страница 78: ...Figure 10 4 Trillium Compact seismometer performance Chapter 10 Performance 16889R10 2020 11 12 Page 78...
Страница 83: ...Figure 12 2 Bottom view Figure 12 3 Side view Chapter 12 Features and Dimensions 16889R10 2020 11 12 Page 83...
Страница 85: ...Figure 12 6 Side view Chapter 12 Features and Dimensions 16889R10 2020 11 12 Page 85...
Страница 95: ...Figure C 1 Cut away illustration of a packed carrying case Appendix C Carrying Case 16889R10 2020 11 12 Page 95...