ADAMII™ ©2016 NanoEnTek Inc.
31
Operation
Measure menu
- Focusing mode
③ Focus option
•
Good Focus
: Clear cell image – sharp cell membrane, dark cell body
with contrast to the bright background
•
Bad focus #1
: Unclear cell image – no contrast to the background
•
Bad focus #2
: Bright and white cell body undistinguishable to the background
•
Bad focus #3
: Blurry and unclear cell image
☞
Note: The example images only relate to CD34 application.
Bright- Bad Focus #2
Bright- Bad Focus #3
Good F
ocus
VS.
Bad F
ocus
[Focusing example – fluorescence channel]
•
Good Focus
: Clear cell image, bright cell body in contrast to the
dark background
•
Bad focus
: Unclear cell image, blurry cell image and donut shaped cell
image
CD45 PerCP- Good Focus
CD34 PE- Good Focus
FITC - Good Focus
CD45 PerCP- Bad Focus
CD34 PE- Bad Focus
FITC - Bad Focus
☞
Note: The example images only relate to CD34 application.
☞
Note: Please check ‘
ADAMII™ Focus guide’
before setting the focus.