NAG/IM/HHP 400/01/22
14
Connect the test instrument to flexible hose/gauge
adaptor and attach to pump through the quick-fit
connection (4).
Turn pressure release valve fully in anti-clockwise
direction.
Generate pressure by squeezing the pump handles
together.
Ensure selector position kept at Low and the handles
are fully squeezed together to achieve up to 30 bar
pressure and then change the selector position to
high for generating required maximum pressure
(depends on model).
Generate the remaining pressure using fine
adjustment valve by turning it in clock-wise direction.
Selection Low / High depends on the range and
easiness of generating of pressure with minimum
strokes.
Release the pressure using the fine adjustment valve
by turning it in anti-clockwise direction and then use
the pressure release valve by turning one turn in
clockwise direction.
After each operation (water medium), the reservoir
must be emptied and let out dry.