NAG/IM/HHP 400/01/22
5
1.
INTRODUCTION
Nagman’s calibration hand pump is designed to manually
generate pressure between 100 / 200 / 350 / 400 bar
(depends on model).
This pump is manufactured with high quality components for
calibrating Analog / Digital gauges by comparison method. It
can be used for lab and field applications.
HHP
is
an ideal pressure source for calibrating pressure
transmitters, pressure transducers and pressure gauges.
There is a Low / High pressure selector switch in the pump,
which is used for selecting low/high pressure output. It is
fitted with a fine Adjustment valve and a Release valve.
The output pressure / vacuum of the hand pump can be
adjusted precisely by the fine Adjustment valve while
calibration. Release valve is used for releasing the pressure /
vacuum after calibration.
The output pressure model ranges are :
Model
Range
HHP 100
Upto 100 bar
HHP 200
Upto 200 bar
HHP 350
Upto 350 bar
HHP 400
Upto 400 bar