NB-205QM/205QMC(Ver. 1)                                                                                                    16/17 

6.3. Temperature Alarm Programming 

To enter this mode, press TEMP SET key in the state of power off and turn power S/W ON. Temperature LED 

will be flashing with a number like 

Ar 1.0

(Factory Programmed deviation is 

±

1

) which means the Tolerance 

Temperature  Range  for  temperature  alarm.  If  temperature  in  display  deviates  more  than  this  programmed 

value for more than 8 minutes, alarm will be activated. 

In this mode, adjust this  value within  adjustable  range(1 ~ 3)  by using up and down  key.  To save  the  new 

value,  press  SET  key.  Then  temperature  display  will  be  turned  to  normal  condition  with  showing  current 

temperature in chamber.   

When  the  incubator  and  alarm  system  are  switched  ON,  or  after  any  value  has  been  reprogrammed,  the 

Temperature Alarm System is inactive until the set point temperature is achieved within 

±

1

  and maintained 

for more than 1 minute in that tolerance range. Alarm is armed through this delay program.   

 

- Over Temperature Cut-Off (visual warning only)     

Thermal  cut-off  have  sensor  attached  on  exterior  surface  of  chamber  seperately  from  main  temperature 

sensor. This protects over-heating when some fail happened in temperature control or sensor.   

Turn the dial of this safety device to be set +5

  from the set temperature point to avoid unnecessary cut-off. 

 

6.4. Shaker Error Messages

   

 

             

                    

 

1) This message “rErrL” represents a low speed error

- Check the platform to ensure that nothing is obstructing its movement. 
- Check the set speed to make sure that it is within the acceptable range(30 to 300rpm). 

Solution

: Use your hand to move platform. While the platform is moving, press and hold the "STOP" and   

"SET" keys simultaneously.                 

 

 

 

 

  

2) This message 

“rErrH” represents a high speed error.   

       - Removal of load or partial load.   

Solution:

 If a vessel has been removed, replace the vessel or restart the machine. 

 

 

 

 

 

 

3) This message “Error” represents an electrical error.   

      - Machine has lost the power during the run. 

Solution:

 Check that the power cord is properly installed and restart. 

Check that the circuit has not been interrupted and restart.   

 

Содержание NB-205QM

Страница 1: ...www n biotek com N BIOTEK NB 205QM QMC Ver 205QMQMC001 1 17 Stackable Incubator Shaker NB 205QM QMC Version 1 OPERATING MANUAL...

Страница 2: ...1 3 Precaution for use 5 2 FEATURE SPECIFICATION 2 1 Features 6 2 2 Specification 6 3 CONFIGURATION CONTROL PANEL 3 1 Configuration 7 3 2 Stacking 8 3 3 Reversible Shaking Motion Platform Optional 9...

Страница 3: ...the faulty caused by user s improper operation excessive use use of incorrect voltage frequency storage in wrong environment mentioned in manual 4 Complete the above table after installation and keep...

Страница 4: ...essed by heavy products When it is damaged it can cause electric leakage or fire Do not handle the power cord with wet hands It can cause an electric shock Do not use the damaged power cord and outlet...

Страница 5: ...incubator To prevent water and experiment material from going into the control panel during the experiment make sure to clean the control panel with a dry towel It can cause a fire and an explosion D...

Страница 6: ...D 4W LED Operating Temperature 5 C 35 C 5 C 35 C Display Digital Display 5 Digit LED Digital Display 5 Digit LED Shaker One Mini Shaker Built in One Mini Shaker Built in Shaking Motion Orbital Orbital...

Страница 7: ...Thermal Cut Off Power Cord Receptacle with Fuse Holder Glass Window Control Panel Screw Tap for Hinge of door for reversible door Shaker Photo 1 Front View Connector of Shaker Vent Holes To cool down...

Страница 8: ...above cover you can find two kits with its screw tighten Install the stacking kit like Photo 4 Turn top cover over and fit it into the frame at top of incubator Photo 5 Locked with Stacking Kit Photo...

Страница 9: ...e Shaking Motion Platform Optional Photo 6 Reciprocating Motion Type Photo 7 Orbital Motion Type 3 4 Shaker Connector Photo 8 Connector Socket of Shaker At Right and Back of Internal Chamber Photo 9 F...

Страница 10: ...MP ON OFF Internal LED Lamp ON OFF Button MUTE Alarm Mute Button SET TEMP Temperature Set up Button Button for Increasing Value in Temperature Setting Mode Button for Decreasing Value in Temperature S...

Страница 11: ...Lamp ON OFF Button MUTE Alarm Mute Button COOL Indication for Peltier Cooling SET TEMP Temperature Set up Button Button for Increasing Value in Temperature Setting Mode Button for Decreasing Value in...

Страница 12: ...t base incubator is level and placed at bottom securely Open top cover of base incubator and find the stacking kits Referring some procedures and photos in page 8 fix the cover to support upper incuba...

Страница 13: ...5 4 6 500ml full 300 3 05 200ml 300 1000 2 3 4 3 85 1000ml full 300 5 15 1000ml full 295 6 5 1000ml full 260 1000 2 3 4 3 10 600ml 300 4 60 600ml 285 4 95 600ml 250 2000 1 3 4 2000ml full 300 2000 1 2...

Страница 14: ...u do not need timer and do need the control for shaking just by On Off switch leave the time set at t 00 00 4 To set the desired mixing speed when shaker stop Press RPM TIME key to switch RPM display...

Страница 15: ...m will automatically stop If mute key is pressed during the shaker alarm 10 minutes of alarm delay will be given If there is door open during RPM alarm shaker will stop and alarm system will be inacti...

Страница 16: ...through this delay program Over Temperature Cut Off visual warning only Thermal cut off have sensor attached on exterior surface of chamber seperately from main temperature sensor This protects over h...

Страница 17: ...f chamber Then lift the shaker up vertically grabbing shaker body and take it out carefully from the chamber Clean the chamber with soft cloths and 70 alcohol After that dry the chamber completely bef...

Отзывы: