6. ALARM AND ERROR MESSAGE 

Press and hold MUTE key for more than 3 seconds until short alarm come out and indicator lamp of alarm is 

turned off. No lighting in alarm indicator means that the alarm system is ON.       

 

6.1. Alarm Event (Factory Programmed) 

1. Door

 : If Door is opened more than 30 second, alarm will be activated.   

- If door is closed when alarming, alarm will automatically stop.   

- If mute button pressed when alarming, alarm will be delayed for 3 minutes.   

- If door is opened continuously after 3 minutes of delay by mute function, alarm will be re-activated. 

2. Temperature :

 

±

1

(Tolerance Range), 8 minutes(Tolerance Time) 

-  If  temperature  is  staying  more  than 

±

1

  from  the  setting  temperature  for  more  than  8  minutes, 

temperature alarm will be activated. (Beep-Beep/ Beep-Beep/~~) 

- If temperature is recovered and stayed in tolerance range for more than 1 minute, alarm will stop.   

- If mute key is pressed during the temperature alarm, 10 minutes of alarm delay will be given.   

3. Shaker

 : 

±

5RPM(Tolerance Range), 60 seconds(Tolerance Time) 

- If RPM value of shaker deviate more than the tolerance range for more than tolerance duration, alarm will 

sound. Based on factory program, if the shaker runs more than 

±

5RPM from the setting RPM for more 

than 60 seconds, audible alarm will be activated with fast flashing on shaker RUN indicator lamp. 

- If RPM recovered into tolerance range, then the alarm will automatically stop.   

-

 If mute key is pressed during the shaker alarm, 10 minutes of alarm delay will be given.   

- If there is door open during RPM alarm, shaker will stop and alarm system will be inactivated. Then, alarm 

system for shaker will be initialized. Once you start the shaker again, alarm system for shaker will be re-

armed by arming delay process as described as below section.     

 

6.2. RPM Alarm Programming 

This mode is to adjust alarm condition such as tolerance time and deviation range.   

To enter RPM alarm programming mode, press RPM SET key and turn Power S/W ON simultaneously. RPM 

display will be flashing with numbers(100, factory programmed). This is the first mode of alarm programming.     

First  Mode

  is  RPM  Alarm  Arming  Delay  Time  which  is  the  length  of  time  that  is  allowed  for  the  shaker  to 

recover after starting the shaker or opening the door, and before the Alarm System is armed again. This helps 

to prevent unnecessary alarms from occurring. Factory programmed value is 100 seconds. For most cases, 100 

seconds are enough time for shaker to achieve set RPM. If it is necessary to adjust, change the value by using 

up and down key. At desired value of time, press SET button to save the value and turn to Second Mode.   

Second Mode

 is Tolerance Deviation Range. It is adjustable between 5 and 10. Allow at least 5 RPM deviation 

to avoid frequent alarming. Adjust the range like previous first mode setting and press SET button at desired 

value to save and to go to Third Mode. 

Third Mode

 is Tolerance Time Delay for higher or lower RPM than tolerance deviation. 60 is recommendable.     

Содержание NB-205QM

Страница 1: ...www n biotek com N BIOTEK NB 205QM QMC Ver 205QMQMC001 1 17 Stackable Incubator Shaker NB 205QM QMC Version 1 OPERATING MANUAL...

Страница 2: ...1 3 Precaution for use 5 2 FEATURE SPECIFICATION 2 1 Features 6 2 2 Specification 6 3 CONFIGURATION CONTROL PANEL 3 1 Configuration 7 3 2 Stacking 8 3 3 Reversible Shaking Motion Platform Optional 9...

Страница 3: ...the faulty caused by user s improper operation excessive use use of incorrect voltage frequency storage in wrong environment mentioned in manual 4 Complete the above table after installation and keep...

Страница 4: ...essed by heavy products When it is damaged it can cause electric leakage or fire Do not handle the power cord with wet hands It can cause an electric shock Do not use the damaged power cord and outlet...

Страница 5: ...incubator To prevent water and experiment material from going into the control panel during the experiment make sure to clean the control panel with a dry towel It can cause a fire and an explosion D...

Страница 6: ...D 4W LED Operating Temperature 5 C 35 C 5 C 35 C Display Digital Display 5 Digit LED Digital Display 5 Digit LED Shaker One Mini Shaker Built in One Mini Shaker Built in Shaking Motion Orbital Orbital...

Страница 7: ...Thermal Cut Off Power Cord Receptacle with Fuse Holder Glass Window Control Panel Screw Tap for Hinge of door for reversible door Shaker Photo 1 Front View Connector of Shaker Vent Holes To cool down...

Страница 8: ...above cover you can find two kits with its screw tighten Install the stacking kit like Photo 4 Turn top cover over and fit it into the frame at top of incubator Photo 5 Locked with Stacking Kit Photo...

Страница 9: ...e Shaking Motion Platform Optional Photo 6 Reciprocating Motion Type Photo 7 Orbital Motion Type 3 4 Shaker Connector Photo 8 Connector Socket of Shaker At Right and Back of Internal Chamber Photo 9 F...

Страница 10: ...MP ON OFF Internal LED Lamp ON OFF Button MUTE Alarm Mute Button SET TEMP Temperature Set up Button Button for Increasing Value in Temperature Setting Mode Button for Decreasing Value in Temperature S...

Страница 11: ...Lamp ON OFF Button MUTE Alarm Mute Button COOL Indication for Peltier Cooling SET TEMP Temperature Set up Button Button for Increasing Value in Temperature Setting Mode Button for Decreasing Value in...

Страница 12: ...t base incubator is level and placed at bottom securely Open top cover of base incubator and find the stacking kits Referring some procedures and photos in page 8 fix the cover to support upper incuba...

Страница 13: ...5 4 6 500ml full 300 3 05 200ml 300 1000 2 3 4 3 85 1000ml full 300 5 15 1000ml full 295 6 5 1000ml full 260 1000 2 3 4 3 10 600ml 300 4 60 600ml 285 4 95 600ml 250 2000 1 3 4 2000ml full 300 2000 1 2...

Страница 14: ...u do not need timer and do need the control for shaking just by On Off switch leave the time set at t 00 00 4 To set the desired mixing speed when shaker stop Press RPM TIME key to switch RPM display...

Страница 15: ...m will automatically stop If mute key is pressed during the shaker alarm 10 minutes of alarm delay will be given If there is door open during RPM alarm shaker will stop and alarm system will be inacti...

Страница 16: ...through this delay program Over Temperature Cut Off visual warning only Thermal cut off have sensor attached on exterior surface of chamber seperately from main temperature sensor This protects over h...

Страница 17: ...f chamber Then lift the shaker up vertically grabbing shaker body and take it out carefully from the chamber Clean the chamber with soft cloths and 70 alcohol After that dry the chamber completely bef...