NB-205QM/205QMC(Ver. 1)                                                                                                    14/17 

5. OPERATION 

1) Place the power switch(located on the right side of the unit) in the "ON" position.   

Display will show current temperature and RPM.     

2) To set the desired temperature, press the "SET" key located underneath the temperature display. 

The set temperature is displayed and will blink continuously. It can be adjusted by using the UP(▲) and 
DOWN(▼) 

buttons

. Once the desired temperature is displayed, press “SET” key to 

save the value. To 

confirm the programmed value, press 

“SET”

 key and confirm the value flashing in display.   

 

Controllable temperature range is as followings: 

NB-205QM : A5

  ~ 60

 

NB-205QMC : 15

  ~ 60

 

 
3) To set the desired time to stop shaker, press and hold the "RPM/TIME" key for approximately 2 seconds 

until LED displays "t 00.00". Press the "SET" key located underneath the RPM/TIME display. The set time 

is displayed and will blink continuously. It can be adjusted by using the UP(▲) and DOWN(▼)

 buttons. 

Once the desired time is displayed, press “SET” key 

again to save. 

    * If you do not press "SET" key to save the desired time, the new set time will not be saved and 

the set time will return to the prior setting. 

** If you do not need timer and do need the control for shaking just by On/Off switch, leave the 

time set at 

t 00.00

 

4) To set the desired mixing speed when shaker stop, Press "RPM/TIME" key to switch RPM display mode 
like r 30 ~ 300. Press the "SET" key located underneath the RPM/TIME display. The set speed is displayed 

and will blink continuously. It can be adjusted by using the UP(▲) and DOWN(▼) 

buttons. Once the 

d

esired speed is displayed, press the “SET” key to 

save. 

5) To change the desired mixing speed, just press and hold 

UP (▲) 

or 

DOWN (▼) keys

 until display value 

reach to desired RPM. Don

t change RPM largely during shaking. Within 100 RPM, adjust RPM as you 

want.   
   

* If you do not press "SET" key to save the desired speed, the new speed will not be saved and the 

set speed will return to the prior setting. 

  ** Speed should never be set below 30 rpm and above 300 rpm. Setting range is 30 ~ 300 rpm.   

***Do not set the shaker in upper stack at more than 200rpm, when using it with stacking option. 

       

Internal Lamp in Chamber 

Press the LAMP key in the control panel to turn ON the internal chamber light.

 

 

Temperature Calibration   

To enter temperature calibration mode, press V key in temperature setting part about 3 seconds. Display will 

be  flashing  with  last  calibrated  value.  In  this  mode,  adjust  display  value  using 

  or 

 

keys  and  press  SET 

button to save the adjusted value.   

 

 

Содержание NB-205QM

Страница 1: ...www n biotek com N BIOTEK NB 205QM QMC Ver 205QMQMC001 1 17 Stackable Incubator Shaker NB 205QM QMC Version 1 OPERATING MANUAL...

Страница 2: ...1 3 Precaution for use 5 2 FEATURE SPECIFICATION 2 1 Features 6 2 2 Specification 6 3 CONFIGURATION CONTROL PANEL 3 1 Configuration 7 3 2 Stacking 8 3 3 Reversible Shaking Motion Platform Optional 9...

Страница 3: ...the faulty caused by user s improper operation excessive use use of incorrect voltage frequency storage in wrong environment mentioned in manual 4 Complete the above table after installation and keep...

Страница 4: ...essed by heavy products When it is damaged it can cause electric leakage or fire Do not handle the power cord with wet hands It can cause an electric shock Do not use the damaged power cord and outlet...

Страница 5: ...incubator To prevent water and experiment material from going into the control panel during the experiment make sure to clean the control panel with a dry towel It can cause a fire and an explosion D...

Страница 6: ...D 4W LED Operating Temperature 5 C 35 C 5 C 35 C Display Digital Display 5 Digit LED Digital Display 5 Digit LED Shaker One Mini Shaker Built in One Mini Shaker Built in Shaking Motion Orbital Orbital...

Страница 7: ...Thermal Cut Off Power Cord Receptacle with Fuse Holder Glass Window Control Panel Screw Tap for Hinge of door for reversible door Shaker Photo 1 Front View Connector of Shaker Vent Holes To cool down...

Страница 8: ...above cover you can find two kits with its screw tighten Install the stacking kit like Photo 4 Turn top cover over and fit it into the frame at top of incubator Photo 5 Locked with Stacking Kit Photo...

Страница 9: ...e Shaking Motion Platform Optional Photo 6 Reciprocating Motion Type Photo 7 Orbital Motion Type 3 4 Shaker Connector Photo 8 Connector Socket of Shaker At Right and Back of Internal Chamber Photo 9 F...

Страница 10: ...MP ON OFF Internal LED Lamp ON OFF Button MUTE Alarm Mute Button SET TEMP Temperature Set up Button Button for Increasing Value in Temperature Setting Mode Button for Decreasing Value in Temperature S...

Страница 11: ...Lamp ON OFF Button MUTE Alarm Mute Button COOL Indication for Peltier Cooling SET TEMP Temperature Set up Button Button for Increasing Value in Temperature Setting Mode Button for Decreasing Value in...

Страница 12: ...t base incubator is level and placed at bottom securely Open top cover of base incubator and find the stacking kits Referring some procedures and photos in page 8 fix the cover to support upper incuba...

Страница 13: ...5 4 6 500ml full 300 3 05 200ml 300 1000 2 3 4 3 85 1000ml full 300 5 15 1000ml full 295 6 5 1000ml full 260 1000 2 3 4 3 10 600ml 300 4 60 600ml 285 4 95 600ml 250 2000 1 3 4 2000ml full 300 2000 1 2...

Страница 14: ...u do not need timer and do need the control for shaking just by On Off switch leave the time set at t 00 00 4 To set the desired mixing speed when shaker stop Press RPM TIME key to switch RPM display...

Страница 15: ...m will automatically stop If mute key is pressed during the shaker alarm 10 minutes of alarm delay will be given If there is door open during RPM alarm shaker will stop and alarm system will be inacti...

Страница 16: ...through this delay program Over Temperature Cut Off visual warning only Thermal cut off have sensor attached on exterior surface of chamber seperately from main temperature sensor This protects over h...

Страница 17: ...f chamber Then lift the shaker up vertically grabbing shaker body and take it out carefully from the chamber Clean the chamber with soft cloths and 70 alcohol After that dry the chamber completely bef...

Отзывы: