77
Trigger timing and
duration of
electrical
stimulation
(for Walk Mode
only)
When the patient's extension/flexion rises to -30°, the device can trigger
electrical stimulation, and when the patient's extension/flexion drops to -5°, the
device's electrical stimulation output is turned off. And the duration of electrical
stimulation is 1s-3s.
Timer:
a) Treatment time: 1-60min, increments of 1min;
b) Timer tolerance: ±2%;
c) When finished, the device can stop output and prompt.
Output Intensity:
Adjustable, 0-100mA, increments of 1mA (at 500Ω load)
Maximum Output
Amplitudes (V):
55 V at 500 Ω Load
Maximum Current
(r.m.s):
50mA r.m.s.at 500 Ω Load
2. Other Specifications
Product Name
MyndStep Foot Drop Stimulator
Model
800102
Содержание 800102
Страница 1: ...1 ...
Страница 2: ......
Страница 36: ...34 No Model Picture Size Connector Remark 1 800111 50mm 50mm square Snap Ф 3 0 mm Standard ...
Страница 37: ...35 2 3 Foot Drop Brace ...
Страница 40: ...38 ...
Страница 85: ...83 ...
Страница 86: ...84 ...
Страница 87: ......
Страница 88: ...MyndTec Inc 800116 Rev 1 0 ...