![MyndStep 800102 Скачать руководство пользователя страница 51](http://html.mh-extra.com/html/myndstep/800102/800102_user-manual_1862692051.webp)
49
4)
Set the parameters
Table: the parameter setting of Walk Mode
Parameter Setting
Range
Default
Increment
Frequency Hz
10-80
50
1
Pulse duration μs
50-500
300
10
Output intensity mA
*
-1
0-100
0
1
*
-1
The output intensity is adjustable only after starting.
5)
Treatment Setup
After setting the output frequency and pulse duration,
click Setup. Adjust the output current and frequency according
to your muscle contraction.
Note: The therapist or patient must ensure the leg side
being treated matches with the left or right leg selection on the
screen, This will enable the selection of the right current
intensity to suit the patient’s needs.