Operations Manual
Page | 30
System Troubleshooting
Symptom
Possible Cause
Remedy
Gel Alarm not
working
FC 03 not adjusted properly
Adjust flow control until gel alarm set correctly
Check RGA V02 (51) for signal – should be off/pulsing
when pumping and on when not pumping
Check RGA V02 output (52) for signal – should be
off/pulsing when pumping and on when not pumping
Flush dwell time
incorrect
FC 05 not adjusted properly
Adjust flow control until flush dwell is set correctly
Flush not working
FC 05 not adjusted properly Adjust flow control until flush dwell is set correctly
Содержание INV2-PAT-7-PRO
Страница 31: ...Operations Manual Page 31 Pneumatic Diagrams ...
Страница 32: ...Operations Manual Page 32 ...
Страница 33: ...Operations Manual Page 33 ...
Страница 35: ...Operations Manual Page 35 ...
Страница 36: ...Operations Manual Page 36 ...
Страница 37: ...Operations Manual Page 37 ...
Страница 38: ...Operations Manual Page 38 ...
Страница 39: ...Operations Manual Page 39 ...
Страница 40: ...Operations Manual Page 40 ...
Страница 41: ...Operations Manual Page 41 ...
Страница 42: ...Operations Manual Page 42 ...
Страница 43: ...Operations Manual Page 43 ...
Страница 44: ...Operations Manual Page 44 ...
Страница 45: ...Operations Manual Page 45 ...
Страница 46: ...Operations Manual Page 46 ...
Страница 47: ...Operations Manual Page 47 ...
Страница 48: ...Operations Manual Page 48 ...
Страница 49: ...Operations Manual Page 49 ...
Страница 50: ...Operations Manual Page 50 ...
Страница 51: ...Operations Manual Page 51 ...
Страница 52: ...Operations Manual Page 52 ...
Страница 53: ...Operations Manual Page 53 ...
Страница 54: ...Operations Manual Page 54 ...
Страница 55: ...Operations Manual Page 55 ...
Страница 56: ...Operations Manual Page 56 ...
Страница 57: ...Operations Manual Page 57 ...
Страница 58: ...Operations Manual Page 58 ...