CHECKING THE INCUBATOR ATMOSPHERE FOR OXYGEN 

The Anaerobic Indicator Solution will indicate traces of Oxygen in the incubator atmosphere by changing 

colour from being clear when the atmosphere is anaerobic to being blue when oxygen is present. The 

Indicator Solution is held in a small bottle located in  the bottom of the incubator and is held in position by 

a plastic holder. The atmosphere from the cabinet is  bubbled through the solution by a small aquatic 

pump. The pump is plugged permanently into an electrical socket. The atmosphere of the incubator can 

be checked for oxygen content by simply switching the socket "ON" on the Control Panel. It is usually 

sufficient to check  the atmosphere perhaps once a day allowing the pump to run for about 2 minutes. If 

the pump is allowed to run continuously, the indicator solution tends to evaporate and requires replacing 

on a regular basis. This is not necessary. 

PREPARING THE ANEROBIC INDICATOR SOLUTION 

Make up the Anaerobic Indicator Solution from the two ingredients provided sufficient to fill the bottle 

about two-thirds full using nine parts of solution "A" and one part of solution "B" 

Now place a cap lightly on the neck of the bottle and place it in a bath of water and bring it to the boil. The 

solution will change from being blue to colourless. Tighten up the cap and remove the bottle from the 

water allow it to cool and pass it into the cabinet. The cap can now be removed and the outlet pipe from 

the pump connected to the bottle. The bottle can now be placed in position inside the bottle holder where 

it will be in full view. 

We recommend replacing the solution on a weekly basis. 

Содержание AW300SG

Страница 1: ...ANAEROBIC WORKSTATION AW300SG MANUAL PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Burnt Mill Elizabeth Way Harlow Essex CM20 2HU UK 44 45 met munro group co uk MUNRO 6 22...

Страница 2: ...T 8 10 CHECKING THE GAS FLOW 9 11 WORKING INSIDE THE CABINET WITH BARE HANDS 10 12 SLEEVE ASSEMBLY 11 13 REMOVAL OF CONDENSATION FROM INSIDE THE INCUBATOR 12 14 THE WATER GAS TAP TUBE 12 15 TRANSFERRI...

Страница 3: ...t Condensation drain bottle Removable exterior port covers complete with retaining clips Bare hand operation ports Catalyst Conditioner Electric supply Gas inlet 6mm push in tube Shelf Bottom door of...

Страница 4: ...CAPACITY 300 Petri Dishes DOOR SEALING Manual GAS Mixed Gas 10 H2 10 CO2 80 N2 OXYGEN REMOVAL FROM AIRLOCK Positive Pressure flushing FABRICATION MATERIAL Acrylic OVERALL DIMENSIONS Length 33 838mm De...

Страница 5: ...XED GAS 35L min The diameter of the pipes connecting the cylinder to the cabinet should be selected according to the length of run to give a minimum pressure drop this will give maximum gas utilisatio...

Страница 6: ...tlet pressure sufficient to maintain pressure at inlet to cabinet Primary gauge showing pressure in bottle 6mm diameter flexible plastic tube 1m long Mount cabinet on strong bench at convenient height...

Страница 7: ...socket Place the conditioning sachet then the catalyst in the channels under the tray in the order shown Conditioner Catalyst Bung Sleeve assy O ring Port Cover Powder Indicator solution bottle fits h...

Страница 8: ...wing the pump to run for about 2 minutes If the pump is allowed to run continuously the indicator solution tends to evaporate and requires replacing on a regular basis This is not necessary PREPARING...

Страница 9: ...chet are in position underneath the shelf The Atmospheric Conditioning Sachet should be put in first to the rear of the cabinet 8 Set the humidistat to 70 R H 9 The commissioning switch will be found...

Страница 10: ...and lift lid as shown e g Gas flow required 35 L minute Pressure Gauge Reading shows 2 Bar Flow Meter Reading 35 35 20 2 L M 1 2 1 73 1 Any adjustment to the gas flow should be made using the flow re...

Страница 11: ...oximity to the left hand port Take hold of the left hand rigid cuff with the right hand and withdraw the left hand Place the left hand rigid cuff over the bung using the right hand Press the cuff over...

Страница 12: ...ollowing the diagrams shown below The assembly consists of four parts 1 The sleeve 2 The rigid cuff 3 The wrist cuff 4 The O rings The assembly fits together as shown in the following sketches 3 Trim...

Страница 13: ...humidistat which is situated inside the unit just behind the transfer lock The humidistat switched the cooling fan on and off automatically as required THE WATER GAS TRAP TUBE The water gas trap tube...

Страница 14: ...e lock and hold in place with the thumb of the left hand while closing the door with the fingers Now remove the top lid of the lock and withdraw the holder Replace the lid and repeat for further holde...

Страница 15: ...re of 160 C for a period of two hours The sachet of conditioner can if necessary be dried at room temperature The chemicals can then be returned to their respective containers for further use When rep...

Страница 16: ...IXED GAS INPUT PIPE SIZE 6 mm DIA INCUBATOR POSITIVE PRESSURE SETTING 6 mm WG INCUBATOR LEAK NOT DETECTED 1 HOUR TEMPERATURE SETTING 36 5 deg C OVER TEMP SETTING OF THERMAL FUSE 98 deg C SPOTLIGHT COM...

Страница 17: ...AW300SG MUNRO Ltd urnt Mill Elizabeth Way Harlow Essex CM20 2HU UK United Kingdom 44 45 Incubator for the environmentally controlled incubation of anaerobic bacteria I certify that the apparatus ident...

Отзывы: