24
2.21.2 Reading Calibration Data
ACTION : Manually record
calibration figures for V1,
V2, V3, I1, I2, I3.
COMMENTS : These
values must be stored with
the serial number of the
M560 as these numbers are
unique to the M560 under
test.
Note Only complete the following operation in 2.21.3 if
the calibration data has been lost and you have recorded
original calibration data.
2.21.3 Entering Calibration Data
ACTION : Click on Manual
Button.
COMMENTS : Enter Value
box is displayed.
ACTION : Enter value .
COMMENTS : Enter
recorded values into V1, V2,
V3, I1, I2, I3 click on O.K.
The Clear button clears all calibration data from the
calibration registers and sets them all to 1.000000.