Setup 

 
Care should be taken with use of the RST Sensor and its cable.  The cable is not designed for 
repeated flexures and should be fixed in a position to minimize unnecessarily sharp bends in 
order to maintain good signal quality and cable life. 
 
First mount the sensor to your fixture for making measurements and connect the sensor cable to 
the back of the controller box.  Connect the host to the RS232 port if it will be used.  Finally 
connect 100V AC power to the power input and then switch the power on. 
 
After power up, the host should wait a minimum of 10 seconds before attempting to communicate 
with the sensor controller.  On power up, the controller display should first clear.  After about 6 
seconds it should show the Multimetrixs logo for a brief moment and then enter into the main 
screen displaying the sample, calibrate, and setup buttons. 
 
 

 
Operation 
 

The MMXI controller can operate “stand-alone” without the connection of an external host to the 
RS232 port.  In this mode of operation, all user input and output is confined to the touch screen 
display on the front of the controller.  However, the controller only displays raw data and therefore 
should be used with a host to convert this raw data into specific units of measurement such as 
film thickness. 
 
In addition to what ever method is used by the user to calibrate and convert raw data to units of 
measure, the sensor controller itself may also perform an internal calibration.   This calibration 
may be made periodically to adjust for system and environment fluctuations or as part of the 
user’s calibration procedure. 
 
Assuming the system has come to a stable temperature, position the sensor head to your 
reference subject and initiate a calibration via RS232 command XCC or the controller’s CALIB 
button on the main screen.  Wait about 15 seconds for successful completion of the calibration.  
Note that the reference you choose may be a known sample similar to the measured subject or it 
could just be free air. 
 
Make readings at this reference position and record them for your own use in converting the raw 
data from the sensor into your specific units of measure. 
 
Note that it is very important to maintain a precise consistent gap between the sensor head and 
the measured subject surface in order to take accurate readings. 
 
The sensor readings are also influenced by the temperature of the system and environment.  The 
system should be allowed to run for at least two hours to reach a stable response.  The more 
stable the environment is, the more stable the readings will be. 
 

 

Содержание MMXI-200

Страница 1: ...Confidential www multimetrixs com Page 1 MMXI 200 Instruction Manual Rev 0 0 Preliminary 6 4 2008...

Страница 2: ...sed on firmware version 1 0B CONTENTS Electrical Connections page 3 Setup page 4 Operation page 4 Touchscreen Display page 5 Serial Command Protocol page 7 Command Set page 7 Configuration commands pa...

Страница 3: ...MMXI controller No other devices should be connected to any part of this port as it could damage the controller and or the connected device Using this port with anything other than the supplied Multi...

Страница 4: ...e should be used with a host to convert this raw data into specific units of measurement such as film thickness In addition to what ever method is used by the user to calibrate and convert raw data to...

Страница 5: ...ection Initial main screen Main screen while sampling on The status information displayed at the bottom of the main screen has the following meaning OK The controller initialized properly ERR The cont...

Страница 6: ...diagnostics unitless G_TEMP Temperature of controller Celsius S_TEMP Temperature inside sensor Celsius STATUS See description of serial command XS 16 bit hex BAUD Host serial port baud rate bps To cha...

Страница 7: ...base 10 variable length text format Integer parameters prefixed with 0x will be parsed as hexadecimal numbers If a command is prefixed with a lowercase letter x the return values if any will be in hex...

Страница 8: ...nd mfal CR Response m CR Configuration loaded F CR Failed CR Syntax failure mfad Description Loads the factory default configuration settings into current working memory Command mfad CR Response m CR...

Страница 9: ...XCS F CR Failed initiation CR Syntax failure XCQ Description Aborts the current calibration prior to completion Command XCQ CR Responses X CR Successfully aborted calibration F CR Failed execution CR...

Страница 10: ...l temperature of the MMXI controller Value is in units of 1 16 degrees Celsius Divide the value by 16 to get Celsius Command XRTi CR Responses X s16 CR Successfully returns the temperature data F CR F...

Страница 11: ...ever use any harsh detergents or solvents on any part of the controller or sensor The MMXI 200 Sensor Controller and the RST Sensor contain no user serviceable parts The user should never attempt to d...

Страница 12: ...ill display text whenever an error occurs When this box is empty as shown above the error log is empty Clicking in the status box will pop up a window displaying the log of errors that have occurred C...

Страница 13: ...ng with a firmware update With the Connected checkbox checked click on the menu item FILE Download Firmware This will start the download process The touchscreen display on the controller will go blank...

Страница 14: ...Confidential www multimetrixs com Page 14 Mechanical Dimensions...

Страница 15: ...Confidential www multimetrixs com Page 15 Notes...

Отзывы: